Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
On some masks there could be a sticky substance on the inside and/or outside of the mask that may manifest itself approximately 6 to 8 months after manufacture of the product. the sticky masks could result in inconvenience and in some cases there is a remote possibility of potential harm to patient.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Vacuette tubes are not filling properly (short fill or no fill).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Manufacturer has already notified accounts plasma samples were being removed from intended use but continued reports of conflicting results when used with other types of blood collection tubes received.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
An issue on the ie33 ultrasound system with the s12-4 transducer. in recent software versions both the acoustic output and patient contact temperature are above specification when the s12-4 is used in cw-mode.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Cobas ampliprep sample racks are out of specification and may cause aborted runs.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Agfa has determined that the current process allowing changes to be made to the impax cv reporting application content via rmat has the potential for misuse. errors may result from uncontrolled changes and these errors could lead to incorrect and clinically significant inaccuracies in the report.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential for failure of growth of haemophilus influenzae. lack of growth would most likely be evident from growth seen on accompanying chocolate plate.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The rf latex in these lots may give a rough negative reaction on some negative specimen.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Collisions and excessive force exerted on the lower portion of the detector arms of the skylight spect systems may have resulted in the detector arms breaking. a collision or excessive force may lead to a crack or fracture in the upper portion of the arm which may cause the detector arms to break and the arm and detector to fall.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Fujifilm has become aware that the epx-2500 operation manual and sales brochure incorrectly identify that the ed-530xt endoscope is capable of being used with the epx- 2500 processor. the operation manual and sales brochure state that when the ed- 530xt endoscope is used with the 2500 processor a "super image" is displayed when it is actually a "standard image" that is displayed.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Microaire surgical instruments has issued a product advisory notice on the pal-900 aspiration tubing due to a labeling error stating "dehp free". although the label contains this error there have been no changes in the material composition of pal-900 aspiration tubing.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Incorrectly gates neutrophil population in patient samples.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
High negative control rates resulting in invalid runs for non-repeating +ve patient specimen runs.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Recent lots of reagents demonstrate a trend towards higher recover of % hba1c and customer inquiries indicate recovery of high values on the cap (college of american pathology) proficiency survey.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential failure of permeate valve could allow feed water into permeate stream.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Autosend not printing patient information in the header.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Product may show poor cd34+ and cd45+ resolution causing the bd stem cell controls to fail resulting in an error message.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Effected lot numbers can produce occassional elevations in sample results of 5-10 ug/ml at 15 ug/ml or 40 ug/ml levels.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Dxc operating software version 5.4 was prematurely installed onto the unicel dxc 600i synchron access clinical analyzer. this software version was intended for all other unicel dxc system configurations but not validated for the dxc 600i system configuration.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Safety advisory with workaround instructions until software upgrade corrects problem.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
This advisory is specific to customers who have upgraded to syngo dynamics version 9.0 with legacy reports generated by syngo dynamics systems version 3.X or earlier. if a legacy report is opened in syngo dynamics 9.0 web portal the report may be deleted and replaced with a new report from the 9.0 reporting system or with a blank report.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Under certqain conditions abnormal a patient burn could occur if attention is not paid to proper patient preparation and room climate conditions.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
In focus release 2.5.0 only when the plan uses an enhanced dynamic wedge (edw) the wedge factor is incorrect under a particular set of circumstances.