Acción
Withdrawal from the market of lot 21080-5H of the "LIFESHIELD Hospira for Plum pump", reference 14240-92-28, manufactured by Hospira Inc., USA, due to a mistake in assembling the punch.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the market of certain batches of the Lifeshield primary equipment for Plum pump, manufactured by Hospira Inc., USA, because they have been manufactured with an incorrect filter in the drip chamber.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that when using GemStar infusion pumps, manufactured by Hospira Inc., USA, the pump will not turn on or off during use.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Update of the System Operation Manual (MOS) and the Technical Manual (MST) of the perfusion systems "Plum A + / A + 3", manufactured by Hospira Inc., USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification of two defects in the software version 15.02 of the Plum 360 infusion pumps, manufactured by Hospira Inc., a Pfizer company, USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the market of certain references and lots of "Nit-Occlud" implantation sheaths, manufactured by pfm medical mepro GmbH, Germany, due to the potential risk of small particles of their soft tip detaching.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal of certain port systems with silicone catheters from the market, manufactured by pfm medical cpp SA, Switzerland, due to the possibility that they may present an irregular surface after high voltage.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the market of certain batches of the GlucoMen® LX Sensor test strips, manufactured by A. Menarini Diagnostics Srl, Italy., Due to the possible obtaining of erroneously high values.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the market of certain batches of the "Relief® ACP (amorphous calcium phosphate) oral care gel or products that contain it", manufactured by Discus Dental LLC, USA, due to the fact that they were filled by mistake with dental peroxide bleaching gel.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Recommendations for the maintenance of uninterrupted power supply systems (UPS) manufactured by third parties, used optionally as certain Computerized Tomography and Nuclear Medicine Systems manufactured by Philips Healthcare,
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that when adjusting the configuration of the Philips IntelliVue MX40 Portable Patient Monitor, product number 865351, manufactured by Philips Healthcare, USA, the ECG alarms are deactivated, preventing them from being sent to the iX monitoring center (PIIC iX).
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that in X-ray systems "DuraDiagnost", with software version 3.0.3, manufactured by Philips Healthcare (Suzhou) Co. Ltd, People's Republic of China, the X-ray image is not acquired correctly due to an anomaly of "firmware" of the detector.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Risk that the acoustic signal emitted by the speaker integrated in the touch monitor model 451220108521 of the X-ray systems "DuraDiagnost" models 712211 and 712214, manufactured by Philips Healthcare (Suzhou) Co. Ltd, China, and indicating the end of the acquisition is not audible.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification that in certain 16-cut MX Model TAC Systems, manufactured by Philips Healthcare Co, Ltd, China, when the image is zoomed and saved in 2D-derived DICOM format, the measurement value of the new image created is incorrect.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification of several software defects in the TAC MX 16-section Systems with version 1.1.4.21426, manufactured by Philips Healthcare Suzhou Co., Ltd., China (formerly Philips and Neusoft Medical Systems Co., China).
Data from Singapore is current through September 2018. All of the data comes from the Health Sciences Authority (HSA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Singapore.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification of several errors in the TAC MX 16-cuts systems, with software version 1.1.4.21425, manufactured by Philips Healthcare, USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that the TAC MX-16 cuts systems, with software version 1.1.3, manufactured by Philips Healthcare, USA, do not show the images when they are used in dental applications in studies with the inclined gantry
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Replacement of the anti-collision sensor and software update of the Philips Allura FC System with configuration ES6600213448, manufactured by Philips India Limited, India, due to the possible failure of the anti-collision system.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
The label of "asymmetric collimators" could be applied incorrectly after switching from a symmetrical collimator to another device with an asymmetric collimator, in Pinnacle3 systems with software version 9.0, manufactured by Philips Medical
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility of the arc of the Diagnostic Trauma System, manufactured by Philips Medical System DMC GmbH, Germany, falling on the table damaging the patient or the user.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility of an incorrect calculation of the dose with the MultiDiagnost Eleva FD system, manufactured by Philips Medical Systems, The Netherlands, due to an error in an algorithm.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Software update in certain HeartStart FR3 Automatic External Defibrillators, manufactured by Philips Medical Systems, USA, due to the detection of two problems.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that the wear of the therapy cable connection of the HeartStart MRx Monitors / Defibrillators, manufactured by Philips Medical Systems, USA, could lead to a failure in the detection of electrode cables / defibrillation blades.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that certain HeartStart MRx monitors / defibrillators, manufactured by Philips Medical Systems, USA, contain defective electronic components on the "PCA" plate