Causa
Coopersurgical is advising all frigitronics cryo-plus cryosugery system users that there are questions regarding the validity of the current high level disinfection procedure for the shaft of the system's hand piece. this notification does not include any cryosurgical tips which currently have a validated high level disinfection procedure. therefore coopersurgical is advising all customers of record to practice all necessary diligence when using the device.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Abbvie has detected an increase in the number of reports of disconnection of the external y-connector from the abbvie 20fr peg tube.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Beckman coulter is making changes to the access intrinsic factor ab instructions for use regarding the concentration of vitamin b12 to the "limitations of the procedure" and "analytical specificiety/interferences".
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
During internal testing zimmer found that a single packaging tray configuration which consists of an inner and an outer sterile tray did not pass the established requirements for seal integrity and seal strength. approximately 2% of either the inner or outer tray did not pass these requirements. this tray configuration is used for the moore hip prosthesis long fenestrated stem.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Avoximeter 1000e instruments may have been sent to customers with the "diagnostic mode" enabled which may lead to potential erroneous results to be displayed.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Evidence that a 26mm uhmwpe insert was isued to this lot code in error should have been 28mm.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
On june 26 2014 one customer complaint was received for this issue (internal ref. # pr162312): a customer reported that a sample had discrepant results when using ssp unitray versus two other typing methods. ssp unitray typed the sample as dpb1*04:01dpb1*81:01. the correct assignment should be dpb1*04:01 dpb1*33:01. this is considered a mistype result.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Nova biomedical has received a report of increased frequency of a no result message being reported when testing.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Table locks may not actuate causing tabletop to unexpectedly float in either the lateral axes or both the longitudinal and lateral axes which could result in patient or operator injury.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The in out & net fluid totals values will not honour the numeric precision configuration which will display a whole number losing decimal point accuracy for low weight patients such as infants.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
A contact issue between the u25 ic and its socket may cause a communication error upon power up (preventing use) when the main and the defibrillator microprocessors are unable to synchronize communications.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Edwards has learned that damage to the crimper's sterile barrier pouch can occur if the crimper while still inside the pouch comes in contact with hard surfaces. this can occur once the pouch is removed from the protective unit carton.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Cleaning recommendations as included in revision g of the directions of use are inconsistent with the 510(k) premarket notification as approved by the fda in 1999: the 510(k) documented the be 183-2 small bore tubing assembly as being for single patient use while the distributed copy states that the tubing assembly is reusable.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Beckman coulter has identified an incompatibility between no foam dxc reagent (ref b64130) and the dxc hydropneumatic parts associated with no foam delivery. dxc no foam reagent is intended to prevent formation of foam and/or bubbles in the dxc waste system of the hydropneumatic subsystems. over time this incompatibility may lead to: breaks in no foam cap assembly (pn a84827) cracks in no foam waste collector (pn a60007/a64665) seepage outside no foam tubing (pn's a10027 a10028 and a67123). this incompatibility does not affect no foam's ability to perform its function and there is no impact to patient results. cracks in the waste collector may stop instrument operation due to vacuum error and potentially lead to biohazardous waste exposure. cracks in the no foam cap and seepage from the no foam tubing may lead to exposure to no foam.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Material hardness on the valve could lead to breakage and failure of the ventilator. dk.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Covidien is issuing a field safety corrective action (fsca) to its puritan bennett 840 ventilator customers to provide new information for the puritan bennett 840 ventilator operator's manual. these new pages provide additional instruction and information for interacting with loss of display events.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Controls may indicate results below specification range when tested with sgpt assay.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Sw update va60b and artis servicepack 1 dfc(m). this correction is a performance issue not related to patient or operator safety.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The product is labelled as 25 mm in length but the files are actually 31 mm in length.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Methodology used to verify calibrator concentration was biased and when re-assayed properly the co2 concetration determined was higher than the value printed on the vial label.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
During recent internal on-board stability studies decreasing calibrator signals were observed for elecsys ft4 reagent kits when stored more than 3 weeks on-board the elecsys systems (cobas e 411 and cobas e 601/602).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Lors de la mise sous blisters de ces dispositifs en salle blanche avant depart en sterilisation une inversion s'est produite entre des pieces d'un lot d'embases tribiales hermes et d'un lot d'embases.