Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
With the intruduction of a replacement mold inconsistency was noted between molds. the previous mold was not producing parts in accordance with the released component drawing & use of new valve block mold resulted in a lower average activation pressure.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Media may allow growth of e. coli atcc25922 which is normally inhibited.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Three units provided to customer with incorrect manufacturer and part number information shown on label. part number provided was hlf-m940-cnv and shows cook urological as the manufacturer (displays "manufactured for:" and is intended for us distribution only). canadian customers should be directed to order identical part number hlf-m940-cnv-can which shows proper manufacturer information for canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Product label text reads 3.2 mm instead of 4.0 mm.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The master list packaged with capture-r ready-id lot id098 contains an error for the donor n2397 typing. the donor n2397 typing should indicate f+ (f-positive).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential problem with the 10-tooth (10t) and 14-tooth (14t) mixer pulley assemblies in the access family of immunoassay systems. it has been determined that in some cases pulley drive components have broken parts or have become loose mis-positioned or disconnected from their drive. when these defects occur the pulley might not turn at all or may turn at an incorrect speed resulting in inadequate mixing of the contents of the reaction vessel which can lead to erroneous test results.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The autospect pro application was only designed to reconstruct cardiac spect data obtained with detectors positioned at 90 degrees or 180 degrees relative to one another. however certain gamma cameras e.G. the marconi axis and irix cameras allow for other relative detector angles. philips has become aware that data acquired at these other angles will not be correctly reconstructed by autospect pro and the results will likely be erroneous.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Issues resulting in application restart issues during start up and logging/diagnostic information and configuration and data presentation issues. (non-real time monitoring/not related to primary monitoring).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
This lot is generating negative control values outside the upper range in the package insert causing an invalid assay run and patient results are not impacted.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
A database deadlock could occur during the image acquisition process affecting the completeness of acquired images. there is the potential that one or more images may be missing from exams. no actual patient injuries have been reported.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Microalbumin multi-calibrator set has transposed vial labels on levels 5 and 6. the concentrations 10 mg/dl and 30 mg/dl listed on the cap labels are correct.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Tempspan cmt may exhibit work time fluctuations and work faster than what the dfu advises.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Customer complaints have been received for falsely elevated patient results that are outside of the customer's expected values (higher than expected some above the dynamic range).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The liaison h. pylori sa kit reports an incorrect volume for the sample diluent component.Kit content states 40ml and specimen diluent component contains 45mls. the component label and instructions for use are correct at 45ml.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Calibrators in lot 182538 do not meet a specification of the manufacturer package insert which is the guaranteed stability of 6 weeks at 2-8 degrees c after reconstitution. stability is reduced to 2 weeks.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The product is being recalled as a result of difficulty encountered when inserting a j-tip guidewire through the distal lumen hub.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The stated tolerance for ctdi(vol) in the instructions for use manual did not accurately represent all systems. on some systems the measured ctdi(vol) value may fall outside of these stated tolerances.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential exists for the spot phototherapy lite light pipe to not hold its set position and sag and expose the patient to possible increased skin surface temperature and reddening of the skin.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Manufacturer identified an issue with dimension? software version 10.2.1. the software version does not permit the user to enter or edit calibration coefficients for applicable assays without support from a siemans customer service representative.