Recalling Organisation: Global Science a Bio-Strategy Company (Bio-Strategy Limited), 241 Bush Road, Albany, Auckland 0632
Causa
Ivds unlicenced by health canada have been exported to new zealand. the manufacturer has initiated a recall of the unlicenced ivds.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Causa
Philips is updating the instructions for use to clarify the user's response to alarms generated by the defibrillator.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Causa
Philips is updating the instructions for use to clarify the user's response to alarms generated by the defibrillator.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
The manufacturer advised of the potential for the brake that holds the gantry in its verical position could release, allowing the gantry to rapidly descend.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
The manufacturer is warning of the potential for an unexpected table drop as a result of collision between the vertical tilt movement and a solid object. after a collision the mechanism may continue to operate, but damage may have been sustained which may lead to the unexpected table drop.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
The manufacturer is warning of the potential for an unexpected table drop as a result of collision between the vertical tilt movement and a solid object. after a collision the mechanism may continue to operate, but damage may have been sustained which may lead to the unexpected table drop.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Varian Medical Systems Australasia Pty Ltd (NZ), C/- BDO Auckland - Christine Koefoed, Level 4, BDO Centre, 4 Graham Street, Auckland 1140
Causa
Varian received reports regarding movement of the blocking washer on the flexible probe with a blocking washer which is a component of the segmented cylinder applicator set. a flexible probe with an unstable or shifted blocking washer must not be used, or a treatment length error may occur.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The silicone suction cup sub-component not being able to confirm the material purity of the suction cup. preliminary testing indicates that some suction cups were manufactured from a non-silicone polymer, polyvinyl chloride (pvc), and that these pvc cups contain latex.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
Medtronic has received reports of dbs leads being damaged at the connector end of the lead when the lead cap is used. the connector end of the lead is the end of the lead connected to the lead extension. tightening or loosening of the setscrew may twist the setscrew connector block and may damage the proximal connector end of the lead.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
Medtronic has received reports of dbs leads being damaged at the connector end of the lead when the lead cap is used. the connector end of the lead is the end of the lead connected to the lead extension. tightening or loosening of the setscrew may twist the setscrew connector block and may damage the proximal connector end of the lead.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
Medtronic has received reports of dbs leads being damaged at the connector end of the lead when the lead cap is used. the connector end of the lead is the end of the lead connected to the lead extension. tightening or loosening of the setscrew may twist the setscrew connector block and may damage the proximal connector end of the lead.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Thermo Fisher Scientific New Zealand Ltd, 244 Bush Road, Albany, AUCKLAND
Causa
The manufacturer advises that this specified lot of oxoid antimicrobial susceptibility discs may contain individual discs that are not impregnated with antibiotic.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Causa
Philips healthcare has identified that it is possible for the philips suresigns vs2+ patient monitors to lock up and stop monitoring during patient use. if the vs2+ is in nbp interval mode with a patient being continuously monitored for spo2 and nbp and the vs2+ locks up, then nbp and spo2 will not be measured and the device alarms will not be functional.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Auckland District Health Board (Health Alliance), CDC Dept Level 4 Health Alliance Building, Auckland City Hospital, Park Road, Grafton, Auckland
Causa
There is potential for the rear portion of the hvad pump's driveline connector housing partially or fully separating from the front portion of the driveline connector. if unattended, the electrical connection to the controller could be affected and loss of mechanical circulatory support may occur.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
The manufacturer has confirmed that with sysmex cs-2000i and cs-2100i systems the last (up to 4) measurements out of a thromborel s vial may be erroneously prolonged (longer pt sec., lowered % of norm, increased inr). this may occur if the reagent in the vial is nearly used up and has been placed on board of the sysmex cs-2000i/cs-2100i systems for more than 4 hours.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Causa
If the philips heartstart xl+ defibrillator/monitor is first powered on and then there is disconnection or an interruption in ac power, the device may be locked out of clinical mode until the user enters the service mode password and performs an operational check.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
Through internal investigation medtronic has identified the potential for uncharacteristically rough metal edges, to occur along the inner diameter of the metal tip side ports of the cannula. to date, medtronic has received no reports of patient injury related to this issue.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Canon Medical Systems ANZ Pty Limited, Building 2, Level 1 Central Park, 666 Great South Road, Penrose, AUCKLAND 1051
Causa
Manufacturer has found an error regarding measurement. this error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/xial plane (rotation around the z-axis), while not being simultaenously rotated around the x-axis and y-axis. manufacturer has found that this orientation adjustement to occur on rare occasions, such as to straighten tilts in head or neck scans, sually with minimal efect on measurements, but sometimes resulting in differences in mength measurement of up to 29% and in volume measurements of up to 50%.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Higher than expected rates of inflammation and/ or endophthalmitis. an extensive review of manufacturing process revealed that some products had trace residual particulates on them. as a precaution manufactureris immediately initiating a recall.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: BioMerieux New Zealand Ltd, Suite 4, The Business Connection Office, 2 Kalmia Street, Auckland 1544
Causa
Under specific conditions the vitek 2 and vitek 2 compact pc systems connected to a laboratory information system (lis) have the potertial to link a test result to an incorrect patient record., the four coditions required to potentially cause an occurence of this issue are:, patient demographics option is enabled, pateint demographics are not associated with the isolate, user does not resolve qualified isolates, laboratory re-uses lab ids (accession numbers) prior to qualified isolates being resolved.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Manufacturer has become aware of discrepancies between the correct measuring ranges provided in the vitros chemistry products instructions for use and those entered on the vitros system for calibration diskettes (cal disks) and assay data disks (add) for these products.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Manufacturer has become aware of discrepancies between the correct measuring ranges provided in the vitros chemistry products instructions for use and those entered on the vitros system for calibration diskettes (cal disks) and assay data disks (add) for these products.