Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by FedEx. 10/14/2005
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
System, X-Ray, Angiographic - Product Code IZI
Causa
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Acción
On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Through post-market surveillance, ela medical has observed that a no-output condition could occur in a limited number of symphony or rhapsody pacemakers. the no-output condition could occur due to metal migration caused by a specific manufacturing process.
Acción
A Dear Doctor letter, dated 10/25/05, was sent to all affected doctors. The letter describes the failures and mentions that pacemaker depend patients implanted with devices manufactured in the first group could require prophylactic replacement.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
System, X-Ray, Angiographic - Product Code IZI
Causa
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Acción
On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Dialyzer, High Permeability With Or Without Sealed Dialysate System - Product Code KDI
Causa
After dialysis machine is disconnected from the patient, blood leakage may occur while unloading the tubing sets from the machine.
Acción
Recall information was forwarded to all Gambro Sale Organizations. Notification was sent to all US consignees on November 2nd and November 3rd, 2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Dialyzer, High Permeability With Or Without Sealed Dialysate System - Product Code KDI
Causa
After dialysis machine is disconnected from the patient, blood leakage may occur while unloading the tubing sets from the machine.
Acción
Recall information was forwarded to all Gambro Sale Organizations. Notification was sent to all US consignees on November 2nd and November 3rd, 2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Introducer, Catheter - Product Code DYB
Causa
Sheath size is larger than dilator size causing insertion difficulties.
Acción
The firm will issue a recall letter to its consignees informing them of the recall. The letter will ask the consignees to return the devices from affected lot numbers for replacement. The returned products will be individually inspected.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Occluder, Patent Ductus, Arteriosus - Product Code MAE
Causa
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (ptfe) plastic lining of the sheath. those shavings could potentially migrate into a patient's bloodstream.
Acción
User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Occluder, Patent Ductus, Arteriosus - Product Code MAE
Causa
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (ptfe) plastic lining of the sheath. those shavings could potentially migrate into a patient's bloodstream.
Acción
User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Occluder, Patent Ductus, Arteriosus - Product Code MAE
Causa
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (ptfe) plastic lining of the sheath. those shavings could potentially migrate into a patient's bloodstream.
Acción
User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Occluder, Patent Ductus, Arteriosus - Product Code MAE
Causa
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (ptfe) plastic lining of the sheath. those shavings could potentially migrate into a patient's bloodstream.
Acción
User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Stents, Drains And Dilators For The Biliary Ducts - Product Code FGE
Causa
The side of the box gives the size of the stent as 10 x 40 instead of the actual size of 7 x 40. the label front is correct.
Acción
Consignees were notified of the initial recall by phone beginning on 10/24/05 and of the extension to the second product on 01/06 /06.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Filler, Bone Void, Calcium Compound - Product Code MQV
Causa
The tissue was collected from donors for whom there is no verifiable identity or consent. the medical records and social histories of the donors cannot be ascertained. the devices which incorporate these donor bone tissues undergo processing, including sterilization, which has been validated to inactivate and/or remove all viral diseases for which human tissue donors are tested.
Acción
Notification sent to all consignees by Fed Ex 10/14/2005.