• Acerca de la base de datos
  • ¿Cómo usar la IMDD?
  • Descargar la data
  • Preguntas frecuentes
  • Créditos
Vista de la lista Vista de las tarjetas
  • Dispositivo 3
  • Fabricante 3
  • Evento 124969
  • Implante 0
Retiro De Equipo (Recall) de Device Recall Welch Allyn ProBP 2400 Digital Blood Pressure Device
  • Tipo de evento
    Recall
  • ID del evento
    76552
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1777-2017
  • Fecha de inicio del evento
    2017-02-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153506
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, measurement, blood-pressure, non-invasive - Product Code DXN
  • Causa
    A defective electrical component in the battery charging circuit in two lots of the probp 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. this potential defect resides with the device, not the battery. the over-voltage condition can result in high nimh battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
  • Acción
    Welch Allyn sent an Urgent Medical Device Recall letter and Response Forms dated February 21, 2017. They were mailed to the first round of customers beginning on Monday February 27, 2017, via US Postal Service Priority Mail. Welch Allyn will follow up with second round of mailings to the additionally identified end users the week of March 20, 2017. All end users will be instructed to confirm that their unit is affected; and if affected, a replacement unit will be provided so that the affected unit can be returned and scrapped. --- Of the 1120 units, a total of 981 units were distributed from Welch Allyn warehouses and 139 units remained under Welch Allyn control and were not distributed. These units have since been destroyed. For further questions, please call (844) 360-8220.
Retiro De Equipo (Recall) de Device Recall MicroAire KWires
  • Tipo de evento
    Recall
  • ID del evento
    76560
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1407-2017
  • Fecha de inicio del evento
    2017-02-21
  • Fecha de publicación del evento
    2017-03-03
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-07-14
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153520
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pin, fixation, smooth - Product Code HTY
  • Causa
    Mislabeling on three lots of k-wires. the incorrect product description " dual trocar" instead of "single trocar" was on the product label on the following parts: 1600-9355ns (lot 61005), 1600-9625ns (lot 60297), and 1604-162ns (lot 62856).
  • Acción
    The firm will mail recall letters to the customers who received the mislabeled k-wires on February 21, 2017. Letters will be sent by UPS to expedite the delivery. The firm will contact the non-responding accounts at least two more times.
Retiro De Equipo (Recall) de Device Recall Guide Wire
  • Tipo de evento
    Recall
  • ID del evento
    76562
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1483-2017
  • Fecha de inicio del evento
    2017-02-17
  • Fecha de publicación del evento
    2017-03-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-19
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153524
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Orthopedic manual surgical instrument - Product Code LXH
  • Causa
    Guide wires sized as .045 x 6 were mis-packaged as .054 x 7 guide wires and those sized as .054 x 7 were mis- packaged as .045 x 6 guide wires. there is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).
  • Acción
    Acumed sent an Urgent Medical Device Recall letter dated 02/16/2017, to all affected customers on 02/17/2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or call to obtain a Return Material Authorization (RMA) number. Customers with questions about recall can call 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada).
Retiro De Equipo (Recall) de Device Recall OEC FlexiView 8800
  • Tipo de evento
    Recall
  • ID del evento
    76568
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1469-2017
  • Fecha de inicio del evento
    2017-02-21
  • Fecha de publicación del evento
    2017-02-27
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-04-24
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153546
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
  • Causa
    Oec flexiview 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
  • Acción
    On February 21, 2017, an Urgent Medical Device Correction letter was sent to one consignee with an affected unit; a field service representative hand delivered the customer communication.
Retiro De Equipo (Recall) de Device Recall Freedom 60 Precision Flow Rate Tubing Sets
  • Tipo de evento
    Recall
  • ID del evento
    76569
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0317-2018
  • Fecha de inicio del evento
    2016-03-10
  • Fecha de publicación del evento
    2018-01-05
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-02-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153551
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pump, infusion - Product Code FRN
  • Causa
    On 2/24/2016 during post sterilization inspection of part number f180 precision flow rate tubing, lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for part number #317036 5 x 6.5 st bag combo. the firm decided to recall precision flow rate tubing and rms high-flo subcutaneous safety needles sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. the firm determined that the issue was large than just the one lot and recalled the product via march 10, 2016 voluntary medical device corrections and removal notification to customers. the recall was conducted without fda notification.
  • Acción
    On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected parties. Customers are encouraged to complete and return the acknowledgement form via fax to: 845-469-5518.
Retiro De Equipo (Recall) de Device Recall HIgHFlo Subcutaneous Safety Needle Sets
  • Tipo de evento
    Recall
  • ID del evento
    76569
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0318-2018
  • Fecha de inicio del evento
    2016-03-10
  • Fecha de publicación del evento
    2018-01-05
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-02-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153552
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Set, administration, intravascular - Product Code FPA
  • Causa
    On 2/24/2016 during post sterilization inspection of part number f180 precision flow rate tubing, lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for part number #317036 5 x 6.5 st bag combo. the firm decided to recall precision flow rate tubing and rms high-flo subcutaneous safety needles sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. the firm determined that the issue was large than just the one lot and recalled the product via march 10, 2016 voluntary medical device corrections and removal notification to customers. the recall was conducted without fda notification.
  • Acción
    On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected parties. Customers are encouraged to complete and return the acknowledgement form via fax to: 845-469-5518.
Retiro De Equipo (Recall) de Device Recall Xcela PICC Kit with ENDEXO and XCELA PASV 5F DL 55CM ...
  • Tipo de evento
    Recall
  • ID del evento
    76570
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0240-2018
  • Fecha de inicio del evento
    2016-11-28
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153553
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter,intravascular,therapeutic,long-term greater than 30 days - Product Code LJS
  • Causa
    Navilyst medical, inc. (nmi) the manufacturer of the xcela picc with pasv, bioflo picc with pasv and the bioflo picc, conducted this recall to the end user level based on information received from greatbatch medical, the manufacturer of the viapeel ptfe peelable introducer. greatbatch medical determined that the products listed in their 11/11/2016 recall notification have the potential for the handles to detach from the sheath during use. nmi has confirmed that the affected sheaths, greatbatch model number 10890-006, had been included in packaged xcela and bioflo picc kits.
  • Acción
    The firm, Navilyst Medical, sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL IMMEDIATE ACTION REQUIRED" letter dated 11/28/2016 by issuing recall notifications to consignees via Federal Express. The letter described the product, problems and actions to be taken. The Consignees were instructed to 1. IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to Navilyst Medical, Inc. (an AngioDynamics Company) o Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. Complete and return the Reply Verification Tracking Form. o If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday  Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. o Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form. o Email Reply Verification Tracking Form (preferred): recall@angiodynamics.com o Fax Reply Verification Tracking Form: Attn: ViaPeel Recall Coordinator Fax number 1-800-782-1357 3. Package and Return the Recalled Product to: AngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, NY 12804, Attn: ViaPeel Recall Coordinator and complete and return the Reply Verification Tracking Form, provided in the recall notification. Attn: ViaPeel Recall Coordinator. If you have any questions, please call: 518-795-1358 or 518-795-1116.
Retiro De Equipo (Recall) de Device Recall PureWick External Catheter
  • Tipo de evento
    Recall
  • ID del evento
    76571
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1485-2017
  • Fecha de inicio del evento
    2016-02-24
  • Fecha de publicación del evento
    2017-03-06
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153554
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Bag, urine collection, leg, for external use, non-sterile - Product Code NNW
  • Causa
    Purewick is recalling external catheter for women wicks because the labeling did not contain the phrase "this product contains dry natural rubber" and/or erroneously included the phrase "latex-free." the white gauze component of the purewick contains dry natural rubber.
  • Acción
    A Urgent Medical Device recall letter dated March 2, 2017 will be sent to customers to inform them that PureWick Corporation is recalling specific lots of PureWick external catheter for women Wicks, Item #625. Customers are informed that PureWick is recalling PureWick Wicks because labeling on product shipped prior to February 22, 2017 either did not contain the phrase This product contains dry natural rubber and/or erroneously included the phrase Latex-free. The white gauze component of the PureWick Wick contains dry natural rubber. PureWick Corporation has received zero complaints or reports of adverse events related to the error in labeling. Customers are informed of the risk to health and the actions to be taken. Customers are instructed to complete the enclosed PureWick Field Correction form and fax to PureWick Corporation at (619) 660-5459 or email to response@purewick.com. Customers with questions or would like assistance completing the PureWick Field Correction Form are instructed to call PureWick toll-free at (844) 584-0734 Monday  Friday from 8:00 a.m. to 4:30 p.m., Pacific Time.
Retiro De Equipo (Recall) de Device Recall Bencox Mirabo PE Liner36/44 Model Number: H1.L61.3644...
  • Tipo de evento
    Recall
  • ID del evento
    76587
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1454-2017
  • Fecha de inicio del evento
    2017-02-27
  • Fecha de publicación del evento
    2017-03-13
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-11
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153590
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
  • Causa
    Lot may contain the wrong instruction (instruction for chinese customer) instead of correct us instructions for use.
  • Acción
    Corentec sent an Medical Device Recall letter dated February 23, 2017. The recalling firm notified the consignee via written notification by email and by phone. The recalled products will be returned to the recalling firm. For further questions please call (512) 327-9997
Retiro De Equipo (Recall) de Device Recall SynchroMed II implantable drug infusion pump
  • Tipo de evento
    Recall
  • ID del evento
    76591
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1694-2017
  • Fecha de inicio del evento
    2017-02-09
  • Estado del evento
    Completed
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153615
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pump, infusion, implanted, programmable - Product Code LKK
  • Causa
    Medtronic received a complaint that there was an error code displayed on the programmer when the physician attempted to interrogate an implanted synchromed ii pump. the error code prevented the physician from updating the pump; however the pump was providing therapy.
  • Acción
    A Medtronic representative visited the clinic on 2/9/2017 and corrected the SynchroMed II pump by resetting the pump memory using a SynchroMed Lab Programmer running on a Medtronic laptop with a validated protocol. Consignee was given a Medtronic "Urgent Medical Device Correction" letter dated February 8, 2017. The letter described the Issue and Actions to be taken by the consignee. For questions contact your Medtronic Representative Medtronic or Neuromodulation Technical Services at 1-800-707-0933.
Retiro De Equipo (Recall) de Device Recall StrataMR Adjustable Valves & Shunts
  • Tipo de evento
    Recall
  • ID del evento
    76595
  • Clase de Riesgo del Evento
    Class 1
  • Número del evento
    Z-1702-2017
  • Fecha de inicio del evento
    2017-02-22
  • Fecha de publicación del evento
    2017-04-07
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-09-19
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153618
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Shunt, central nervous system and components - Product Code JXG
  • Causa
    Incorrect valve rotor orientation due to misalignment and/or off-center positioning of the stratamr locator tool: when the stratamr valve is set in the incorrect orientation, higher valve opening pressures result leading to a corresponding underdrainage in csf fluid performance. this condition may result in the following adverse health consequences: ventriculomegaly, headaches, nausea, vomiting and lethargy. if left untreated, underdrainage has the potential to lead to coma and death.
  • Acción
    The firm has initiated a CAPA and issued notification via letter to US Customers on February 22, 2017. As of the date of the report (2/22/17) the firm was still in process of notifying foreign customers. Recipients are instructed to: 1) Cease use of the listed product, 2) Account for number of product units implanted or discarded and return any unused units by completing an attached individualized customer product accountability form provided in the letter, 3) If any of the indicated products have been implanted in patients, to adjust the valves according to instructions listed in an included attachment. Recipients are asked to fax a customer acknowledgment response form or scan/take a picture of the form and email to rs.mnsfca@medtronic.com with "StrataMR Recall" in the subject line.
Retiro De Equipo (Recall) de Device Recall 7F Prelude Short Sheath Introducer
  • Tipo de evento
    Recall
  • ID del evento
    76596
  • Clase de Riesgo del Evento
    Class 1
  • Número del evento
    Z-1497-2017
  • Fecha de inicio del evento
    2017-02-16
  • Fecha de publicación del evento
    2017-03-24
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-08-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153619
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Introducer, catheter - Product Code DYB
  • Causa
    Merit medical systems, inc. announces a voluntary field action for sheaths included with the 7f prelude¿ short sheath introducers because the sheath marker tips may detach during use.
  • Acción
    Consignees notified on 2/16/17 to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact your Merit Sales Representative or Merit Customer Service at 801.208.4381.
Retiro De Equipo (Recall) de Device Recall HeartWare Ventricular Assist System
  • Tipo de evento
    Recall
  • ID del evento
    76600
  • Clase de Riesgo del Evento
    Class 1
  • Número del evento
    Z-1538-2017
  • Fecha de inicio del evento
    2017-02-03
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153628
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Ventricular (assisst) bypass - Product Code DSQ
  • Causa
    Product recall of all hvad controllers currently on the market and to exchange them for the next generation hvad controller (controller2.0).
  • Acción
    The firm will release a communication to their consignees stating the following: As a reminder, as with all HVAD Controllers, continue to reinforce the following with your patients and staff at all opportunities: " Patients should continue to have a backup HVAD Controller ready at all times in the event of a primary HVAD Controller failure. " Staff only: The driveline extension cable is to be used during the pre-implant test only. It is not intended to be used after the pump is implanted in the patient. Hospital and Clinician Actions (to be executed in the following order): 1) Review the enclosed notice and forms, and forward the notice to those individuals within your organization who need to be aware of its contents. 2) Complete, sign, and return the Acknowledgement Form to HeartWare within thirty (30) days of receipt of this letter. 4) Quarantine and replace affected HVAD Controllers, DC Adapters, Instructions for Use, Emergency Responder Guides and Patient Manuals in hospital inventory after training is complete. For every patient, quarantine and replace the following under clinical supervision in an environment where appropriate support equipment is readily available: - Primary and Backup HVAD Controller; - Affected DC Adapters; and - Patient Manual and Emergency Responder Guide. Clinicians are reminded not to perform an HVAD Controller exchange during an active electrical fault alarm as the HVAD Pump will be running a single stator. If an electrical fault is present, download patient log files and contact your HeartWare representative to resolve the electrical fault before executing the controller exchange. 5) Return all quarantined HVAD Controllers and DC Adapters to HeartWare. Your HeartWare representative will assist you with this process. 6) Completion Form. Once affected product in inventory has been identified and returned, complete and return the attached Completion Form to CON2.0@heartware.com or your HeartWare representative no late
Retiro De Equipo (Recall) de Device Recall Endurant / Endurant II 23mm and 25mm Bifurcated Stent...
  • Tipo de evento
    Recall
  • ID del evento
    76593
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1543-2017
  • Fecha de inicio del evento
    2017-02-27
  • Fecha de publicación del evento
    2017-03-15
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-03-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153631
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, endovascular graft, aortic aneurysm treatment - Product Code MIH
  • Causa
    Medtronic is initiating a voluntary urgent medical device recall for a subset of endurant/ endurant ii bifurcated stent graft systems of specific models and serial numbers as this specific subset of stent grafts has greater susceptibility to fabric permeability variations that may be associated with endoleaks observed during the initial implant procedure.
  • Acción
    Medtronic sent an Urgent Medical Device Recall letter dated March 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate that your facility has received a potentially affected Endurant /Endurant II 23mm and / or 25mm Bifurcated Stent Graft System from the identified subset. As a result, Medtronic is asking that you take the following actions: Identify and quarantine unused Endurant /Endurant II 23mm and 25mm Bifurcated Stent Graft System from the identified subset of models and serial numbers that are in your inventory . Return unused products from the attached model and serial number list that are in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570 (#4) and referencing this communication to initiate a return and credit of unused product. Your Medtronic sales representative can assist you in the return of this product as necessary. Complete the attached Customer Confirmation Certificate and email it to RS.CFQFCA@Medtronic.com or fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. For further questions please call Customer Service at 1 (800) 633-8766.
Retiro De Equipo (Recall) de Device Recall Mammotome Revolve Dual Vacuum Assist Biopsy System
  • Tipo de evento
    Recall
  • ID del evento
    76603
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1512-2017
  • Fecha de inicio del evento
    2017-02-28
  • Fecha de publicación del evento
    2017-03-08
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-08-03
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153632
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Instrument, biopsy - Product Code KNW
  • Causa
    The device did not have revolve stereo probe labeling, no product code indicating the size and length of the probe, no sterilization sticker or lot number.
  • Acción
    Devicor sent an Urgent Medical Device Recall letter dated March 10, 2017 to all affected consignees via first class mail. The recall response rate will be tracked and reconciled by consignee's response form, certifications of product inventory, and/or disposal. Consignees were asked to complete and return the enclosed response form as soon as possible. Consignees with any questions were instructed to call Mammotome Customer Service at 1-877-926-2666. For questions regarding this recall call 513-864-9158.
Retiro De Equipo (Recall) de Device Recall MEDHOT MTI 2000 THERMAL IMAGING SYSTEM
  • Tipo de evento
    Recall
  • ID del evento
    76622
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2057-2017
  • Fecha de inicio del evento
    2017-01-14
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153694
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, telethermographic (adjunctive use) - Product Code LHQ
  • Causa
    Indications for use are not cleared or approved for market in the u.S.
  • Acción
    Correct software to eliminate any reference to temperature. Test function and notify users of software update with new user instructions.
Retiro De Equipo (Recall) de Device Recall Covidien Curity Wet Dressing
  • Tipo de evento
    Recall
  • ID del evento
    76606
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1742-2017
  • Fecha de inicio del evento
    2017-03-03
  • Fecha de publicación del evento
    2017-03-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153639
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Dressing, wound and burn, hydrogel w/drug and/or biologic - Product Code MGQ
  • Causa
    Product sterility is compromised due to breach of sterile barrier.
  • Acción
    Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall Covidien Kerlix Super Sponge Saline Dressing
  • Tipo de evento
    Recall
  • ID del evento
    76606
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1743-2017
  • Fecha de inicio del evento
    2017-03-03
  • Fecha de publicación del evento
    2017-03-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153640
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Dressing,wound,occlusive - Product Code NAD
  • Causa
    Product sterility is compromised due to breach of sterile barrier.
  • Acción
    Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall Curity Sodium Chloride Dressing
  • Tipo de evento
    Recall
  • ID del evento
    76606
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1744-2017
  • Fecha de inicio del evento
    2017-03-03
  • Fecha de publicación del evento
    2017-03-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153641
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Dressing,wound,occlusive - Product Code NAD
  • Causa
    Product sterility is compromised due to breach of sterile barrier.
  • Acción
    Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall Curity sodium chloride dressing
  • Tipo de evento
    Recall
  • ID del evento
    76606
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1745-2017
  • Fecha de inicio del evento
    2017-03-03
  • Fecha de publicación del evento
    2017-03-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153642
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Dressing,wound,occlusive - Product Code NAD
  • Causa
    Product sterility is compromised due to breach of sterile barrier.
  • Acción
    Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall AMS636, 60" Microbore extension set.
  • Tipo de evento
    Recall
  • ID del evento
    76607
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1484-2017
  • Fecha de inicio del evento
    2017-02-20
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153644
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Set, administration, intravascular - Product Code FPA
  • Causa
    Incorrect priming volume printed on the device package.
  • Acción
    Recall letters were sent to the customer via email and fax on 2/20/2017. They were asked to take the following actions: The recall will be conducted at no cost to you. To effect return of the recalled product and replacement with new inventory, we request your cooperation in the following steps: 1. Check your stock for the products included within the scope of the recall. Immediately cease use and distribution, and quarantine all affected product immediately; 2. Count and document your affected inventory on the "Recall Acknowledgement and Inventory Return Form; 3. Fax or email the form to Vygon including customer contact information; 4. Vygon customer service will contact the customer and issue a Return Merchandise Authorization number; 5. Return the product as well as the original signed Recall Acknowledgement and Inventory Return Form using the instructions from Customer Service involving product pick-up; 6. All inventory will be credited or replaced at no cost to you. If you have questions, please contact your local sales representative or Vygon's Customer Service Department at 1-800-4 73-5414, or by e-mail at customerservice@vygonus.com.
Retiro De Equipo (Recall) de Device Recall Altius MINI OCT Posterior Spinal Fixation System
  • Tipo de evento
    Recall
  • ID del evento
    76610
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1691-2017
  • Fecha de inicio del evento
    2017-03-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-07-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153660
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Intervertebral fusion device with bone graft, lumbar - Product Code MAX
  • Causa
    Zimmer biomet is recalling the altius m-ini oct posterior spinal fixation system because the driver is unable to mate with the screw head tulip.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall letter on March 20, 2017, to all affected consignees. The letter identified the product, the problem and the action to be taken by the consignee. Consignee's Responsibilities 1. Review this notification and ensure affected team members are aware of the contents. 2. Prior to sending Altius Kits to central processing, perform a functional check of the driver withall screws in the kit to ensure they mate. If the drivers do not mate, please return in accordance with Step 3b. 3. Complete the Certification of Acknowledgement portion of Attachment 1  Inventory Return Certification Form  Phase I. b. If the driver is unable to mate with the screws in the set, immediately sterilize and return the affected nonconforming product from your distributorship and affected hospitals within your territory along with a completed Attachment 1  Inventory Return Certification Form  Phase I and Attachment 2  Certificate of Sterilization to Zimmer Biomet. i. For each return, send a copy of Attachment 1 to jeremy.hansmann@zimmerbiomet.com ii. Include a hardcopy of Attachment 1 with your shipment for immediate processing. iii. Include a copy of Attachment 2  Certificate of Sterilization with returned instruments. iv. Mark the outside of the returns box(es) clearly with RECALL. 4. Distributors will notify their accounts to which the product was further distributed and to locate and return products from these accounts. Using the Additional Accounts Form provided with the email notice sent to your facility, return contact information for any additional hospitals that have the affected product. If there are no additional users to notify, please confirm this by annotating and returning the form provided indicating that no additional users have been identified. 5. Retain a copy of your field action acknowledgement and product return forms for your records in the event of a compliance audit of your facility. 6. If
Retiro De Equipo (Recall) de Device Recall Artis Q and Q.Zen Fluoroscopic xray system
  • Tipo de evento
    Recall
  • ID del evento
    76615
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1466-2017
  • Fecha de inicio del evento
    2017-02-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153674
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Interventional fluoroscopic x-ray system - Product Code OWB
  • Causa
    A gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. such liquids may include disinfectants, cleaning agents and/ or bodily fluids. in rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch.
  • Acción
    A Customer Safety Advisory Notice (CSAN) was mailed to affected customers on 02/27/2017 to inform them of the issue. The notice provided information on what actions to take should the foot switch fail to function, the risks involved and what corrective actions they are taking to fix the issue.
Retiro De Equipo (Recall) de Device Recall Artis Zee and Zeego Fluoroscopic xray system
  • Tipo de evento
    Recall
  • ID del evento
    76615
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1467-2017
  • Fecha de inicio del evento
    2017-02-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153675
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, x-ray, angiographic - Product Code IZI
  • Causa
    A gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. such liquids may include disinfectants, cleaning agents and/ or bodily fluids. in rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch.
  • Acción
    A Customer Safety Advisory Notice (CSAN) was mailed to affected customers on 02/27/2017 to inform them of the issue. The notice provided information on what actions to take should the foot switch fail to function, the risks involved and what corrective actions they are taking to fix the issue.
Retiro De Equipo (Recall) de Device Recall Axiom Artis Fluoroscopic xray system
  • Tipo de evento
    Recall
  • ID del evento
    76615
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1468-2017
  • Fecha de inicio del evento
    2017-02-27
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=153676
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Interventional fluoroscopic x-ray system - Product Code OWB
  • Causa
    A gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. such liquids may include disinfectants, cleaning agents and/ or bodily fluids. in rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch.
  • Acción
    A Customer Safety Advisory Notice (CSAN) was mailed to affected customers on 02/27/2017 to inform them of the issue. The notice provided information on what actions to take should the foot switch fail to function, the risks involved and what corrective actions they are taking to fix the issue.
  • First
  • Previous
  • 1
  • 2
  • 3
  • 4
  • 5
  • Truncate
  • Next
  • Last

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.

Descargar la data

La International Medical Devices Database está bajo la licencia Open Database License y sus contenidos bajo la licencia Creative Commons Attribution-ShareAlike . Al usar esta data, siempre citar al International Consortium of Investigative Journalists. Puede descargar acá una copia de la base de datos.

Descargar todo (zipped)