Causa
Contamintaion of the mirror from the optical system can lead to false positive signals in the early phase of measurement which can lead to the following failure pattern on a cobas h 232: "tnt elevated" displayed during the measurement but the final results is trop t (less than) 50ng/l iqc level-ii fails e-402 during a measurement.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Internal testing showed that the nichols advantage acth cartridge lot # 62-404296 may not meet claims in the directional insert concerning correlation with the irma acth assay catalg # 40-2194.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Several complaints of false positive with the captia varicella-zoster virus (vzv) igm elisa.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
In artida ultrasound systems under certain conditions the operation panel can become unresponsive (a "lockup").
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The onboard stability for the elecsys ft3 reagent will be reduced from 8 weeks to 4 weeks on elecsys 2010 modular analytics e and cobas e601 systems.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
When sample couch parameter files are edited the densities altered and "save" is selected the sample files will be updated with these new user defined densities. subsequently when a software upgrade is loaded under the conditions described below the modified couch densities in these sample files will be overwritten with the default value of 1.000.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Siemens has become aware that it is possible for patients or operators to have their fingers trapped and pinched between the table top and the bottom frame of the 550 txt treatment table which is part of the affected systems listed in section 64(a). secondly and unrelated siemens received a report on an incident in which a cable inside of the 550 txt treatment table had been damaged due to contacting a moving part. the resulting short-circuit caused overheating of an electronic circuit board inside the table base.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
An internal metal shield for emi protection may short rendering the unit inoperable with audible alarm to alert caregivers or may create a spark in an oxygen rich environment if a double fault existed.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The quality control criteria for the kit are not met. the results of test runs are invalid. there is no risk of a false positive.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Customers shipped the paceart system - get connected first edition may have received a cd that contains the export module software but is mislabeled as ttm management sound card enabled software.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Symbiq failure to detect air in line due to a liquid droplet randomly adhering to the inner wall of a symbiq microbore administration set.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Mit is making changes to the med-jet label and instruction for use. the intended use of the device were clarified as requested by health canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Steris has learned through customer feedback that vision sc washers are experiencing nuisance alarms and/or aborted cycles that are adversely affecting the reliability and operation of their unit.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
This lot of media was not manufactured to the specified formualtion which resulted in an incorrect ratio of bile salts being added. continued use may result in false negative reporting incorrect bacteria counts and/or incorrect bacteria identification.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The wire guide in the fusion omni-tome pre loaded sphincterotome being loaded incorrectly into the device resulting in the incorrect end of the wire guide exiting the sphincterotome. the preloaded wire guide has a black floppy distal tip which could be located at the proximal/user end and leaving the stiff end of the wire guide at the distal/patient end of the sphincterotome. difficulty advancing the wire guide may be experienced.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
During biopsy procedure patient was scanned with routine helical protocol. although the image was acquired no real-time image was displayed.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
If the "insufficient reagent" or "sample probe sampling error" alarm is displayed during an analysis the very next test results will be erroneous.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Substrate is exhibiting a blue color which may cause false positive results.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Software deficiency in the masking of laser pulses to a portion of the cornea.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Tubing leading from the pump adapter to the distal end of the product became detached rendering the set unusable.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
Notas adicionales en la data
Causa
Possible interruption of therapy may occur when an e321 error code occurs while using a plum a+/a+3 infuser. the e321 error code occurs when the plum a+/a+3 infuser is operating on ac power and the software detects that the battery could not be fully recharged within 8 hours.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
It was found that there was a positive shift >10% at the upper limit of the normal reference range for the immulite systems free t3 method as the values do not align with the patient clinical picture when interpreted with other thyroid markers.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Manufacturer has confirmed the potential for over-recovery of the c-peptide cvm top level with cvm lots 012 (level j) and 0101 (level 4) when used with the immulite systems c-peptide assays. in some instances this may lead to results outside the range published in the cvm instructions for use (ifu). internal testing demonstrated that all other levels for cvm lots 012 and 0101 are unaffected. the potential over-recovery is limited to the top level cvm and patient results are not impacted.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Decrease in the concentration of a raw material during manufacturing which could affect long term performance of the product.