Causa
When patient permanent blood type is processed from lars the permanent blood type is updated but the comments are removed. dk.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
An issue was identified with the use of auto-fetch feature of the pacs-iw viewer. with auto fetch in use another patient's image may appear without user intention.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The ph is at 6.2 to 6.3 versus a specification of 6.5 to 6.9. the material contains a ph indicator (bromcresol purple) which is used as part of the mechanism of differentiation. the low ph is causing organisms that should give alkaline reactions to not give the correct reaction two complaints originating in the us (dec 2010 and june 2011) have been filed with the manufacturer in regards to failed ph results. both facilities properly prepared the medium and both found inconsistent performance based on quality control specifications. however proper qc was performed leading the facilities to not use the product.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Software error may permit the chair to recline too far possibly causing the user to slide out backwards head first.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The advisory notice is associated with the visx excimer laser system (licence 2145). this action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure in the ac power distribution printed circuit board (pcb) in the visx excimer laser system specifically models star s2 star s3 star s4 custom vss and star s4 ir.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Medtronic is voluntarily recalling specific production lots of the covidien endo peanut blunt dissector 5mm. this voluntary recall is being conducted following reports that the cotton tip of the device may disengage due to insufficient adhesive application during the manufacturing process.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Software update to prevent unwanted travel of phs.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Can create a situation where a pixel-by-pixel heterogeneity correction is used in the dose calcuation but the system output indicates assigned densities have been used (the reverse is also true).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Physio-control has become aware of an issue whereby the lifepak cr plus aed or lifepak expressaed may fail to initiate voice prompts when the on/off button is pressed and the lid is opened due to an internal component (reed switch) that can intermittently become fixed in the closed position. physio-control has determined that there are presently 25178 devices in the field of which approximately 9600 may have this faulty component. a defibrillator in this condition is not delivered and a patient is not resuscitated. all reports received from customers have been noted during initial set up and testing of the device.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
If the dicom coordinates option is turned on in the reports pull down menu the following issues are seen on the resulting print out issue 1 the coordinates will be in dicom space thus not matching the setup reference dialog. issue 2 the shift coordinates are reported incorrectly.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The unit carton of the label incorrectly states that the amount of gentamicin in the bone cement is 1.0 m.I.U. (million international units). the actual amount of gentamicin in the bone cement is 0.5 m.I.U. the product mas made to specification and the information is correct on the instructions for use (ifu) and the foil pouch. the only incorrect artwork is on the carton label.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Synchron cx5pro level sense boards have incorrect capacitor. boards my fail prematurely which would result in level sense error messages.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The manufacturer has been made aware of an applicaton freeze occuring during treatment initiation. when a forced logoff occurs while the dummy wire is extended during a treatment session the partial fraction generated by the system will be incorrect.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The hematology module is generating test data for tests that have not been run. when the system reports an automated differential the system will show on the final report (cumulative) all the diff. parameters.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Advisory letter - patient may be subject to contrast agent injection without images being obtained.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The whole-body hyperthermia device ht-2000 was sold into canada without holding a valid medical device licence.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Smith & nephew has found that specific lots of the cannulated reusable obturator will not fit the disposable clear-trac cannula identified for use with the product.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Siemens has confirmed customer complaints regarding low recovery of the tacrolimus flex reagent cartridge df107.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Two conditions have been identified with the philips medical systems brilliance big bore version 2.2.1 tumour localization that may lead the user to misinterpret the displayed data.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Sw errors between impac multi-access or mosaic systems & obi versions 1.0.15 & obi 1.2.05 together with a 3rd party treatment planning system like adac pinnacle or cms xio can cause patient positioning errors.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The reading of cpap pressure within the software to which the embletta gold was connected showed a reading of approximately 2cm h2o lower than that of the cpap.