Retiro De Equipo (Recall) de Maquet/Datascope Intra-Aortic Balloon Pump (IABP)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Maquet Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-00834-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2017-07-04
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Getinge has identified a potential issue with the solenoid driver board within the unit of the cs100i, cs100 or cs300 iabp. a complaint was received involving a cs300 iabp that did not pump due to an electrical test failure code #58 (power up vent tests fail), maintenance code #3, and an autofill failure, which has been associated to a patient death due to the failure of the device to initiate therapy.An electrical test failure code #58 is caused by a solenoid valve requiring more power than the solenoid driver board can deliver to open the valve. the lack of power prevents the coil from moving the plunger causing the valve not to open.These issues could result in an interruption and/or delay in therapy to the patient during use and/or prior to using their cs100i, cs100 or cs300 iabp. a sudden interruption of therapy could result in unsafe, haemodynamic instability. this issue also applies to any system 98 or system 98xt iabp which was converted to a cs100i or cs300 iabp.
  • Acción
    Getinge advises: 1. Pursuant to the WARNINGS section of the Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy; 2. Until the service is performed, Getinge recommends powering on the IABP prior to inserting the IAB catheter to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform; 3. In the event the IABP fails to complete the self-test and exhibits electrical test failure code 58, remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office; 4. In the unlikely event that a sudden interruption of therapy occurs, transfer the patient to an alternative IABP. If an alternative IABP is unavailable; manually inflate the IAB with air or helium and immediately aspirate. Please refer to the intra-aortic balloon catheter instructions for use; and 5. Each affected facility will be contacted by Maquet/Getinge Service Team to schedule on-site service.

Device

  • Modelo / Serial
    Maquet/Datascope Intra-Aortic Balloon Pump (IABP)CS100i IABPPart Numbers: 0998-UC-0446HXX, 0998-UC-0479HXXCS100 IABPPart Numbers: 0998-00-3013-XX, 0998-UC-3013-XXCS300 IABPPart Numbers: 0998-00-3023-XX, 0998-UC-3023-XXDistributed between 24 March 2003 through to 11 December 2013ARTG Number: 118266