Retiro De Equipo (Recall) de Active Intact PTH

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Beckman Coulter Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    56250
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0585-2011
  • Fecha de inicio del evento
    2010-04-22
  • Fecha de publicación del evento
    2010-12-09
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2012-05-07
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Radioimmunoassay, parathyroid hormone - Product Code CEW
  • Causa
    The recall was initiated because beckman coulter has confirmed that the stated stability of the dsl-10-8000 i-pth elisa and dsl-8000 i-pth irma assays may not be achieved for the lot numbers affected by the recall. the affected i-pth elisa and the i-pth irma kit lots may produce falsely elevated patient results. internal stability testing showed an average 16% elevation for i-pth elisa lot 99.
  • Acción
    Beckman Coulter sent an Urgent Product Corrective Action Letter to their consignees dated June 18, 2010. The letter identified the product, the problem, and the action to be taken. Consignees were instructed to: (1) Discontinue use of all materials provided with I-PTH ELISA kit lot number 992751 and I-PTH IRMA lot numbers 090013, 090121, and 090372. (2) Review historical external quality control results to ensure the assay performance of these lots was within laboratory specifications. Retesting may be necessary if their laboratory uses only the Control materials provided with the I-PTH ELISA and IRMA kits. (3) Abnormal elevated patient results that are obtained with these lot numbers and that are inconsistent with the patient's clinical status should be reviewed in the context of other clinical diagnostic tests. (4) I-PTH ELISA kit lot number 992751 and I-PTH IRMA lot numbers 090013, 090121, and 090372 will no longer be shipped. Due to the calibrator instability recently encountered with DSL-10-8000 I-PTH ELISA and DSL-8000 I-PTH IRMA assays, Beckman Coulter has decided to immediately discontinue these two part numbers. Consignees were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Complete and return the enclosed response form within 10 days so Beckman Coulter can be assured that they received this important notification. If they have any unused product the were asked to contact the numbers listed below: (1) In the United States, contact Customer Service at 1-800-526-382.1 (2) Outside of the United States, consignees were asked to contact their local Beckman Coulter Representative Consignees were told to contact DSL Technical Support Center at 1-800-231-7970 in the United States, or their local Beckman Coulter Representative with any questions. For questions regarding this recall call (714) 961-3634.

Device

  • Modelo / Serial
    Lot Numbers: 090013, 090121, and 090372
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - USA including MD, NY, GA, NJ, CA, MI, TX, PA, IA, and IN. Internationally to Austria, Brazil, China, Colombia, Czech Republic, France, Germany, Hong Kong, India, Japan, Mexico, and the Russian Federation
  • Descripción del producto
    Active INTACT PTH, Part Number: DSL-8000 || The DSL-80000 ACTIVE Intact PTH IRMA Kit provides materials for the quantitative measurement of Intact PTH (I-PTH) in serum or plasma. The assay is intended for in vitro diagnostic use.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Beckman Coulter Inc., 200 S Kraemer Blvd, Brea CA 92821-6208
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA