Retiro De Equipo (Recall) de ARCHITECT STAT Myoglobin Calibrators

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Abbott Laboratories MPG.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    33845
  • Clase de Riesgo del Evento
    Class 3
  • Número del evento
    Z-0400-06
  • Fecha de inicio del evento
    2005-10-17
  • Fecha de publicación del evento
    2006-01-18
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2006-05-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Myoglobin, Antigen, Antiserum, Control - Product Code DDR
  • Causa
    Some of the kits were shipped at the incorrect shipping temperature. they were shipped ambient, but should have been shipped frozen.
  • Acción
    Abbott telephoned the laboratory end users on 10/17/05, informing them that Abbott discovered that ARCHITECT STAT Myoglobin Calibrators, lot V96128, were shipped to their facilities using the incorrect shipping conditions between August 29th and September 20th 2005. The kits should have been shipped frozen, but were shipped ambient. The error has been corrected. All shipments made after September 21, 2005 have been shipped correctly. Any data generated to date using this kit of calibrators is acceptable, providing that a valid calibration was achieved and product labeling was followed. However, long term use of this kit may result in calibration failures or controls out of range. The accounts were requested to mark the kit(s) of this lot currently in inventory to distinguish them from the replacement kit(s) Abbott will be sending to them. As a precaution, Abbott requested the labs to discontinue use of these kits and destroy them the replacement kits are received.

Device

  • Modelo / Serial
    list 02K43-01, lot number V96128
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Texas, Washington, Montana, Pennsylvania and Kentucky, and internationally to Korea
  • Descripción del producto
    ARCHITECT STAT Myoglobin Calibrators; list 02K43-01; each kit contains 6 bottles of calibrators ranging from 0 ng/mL to 1200 ng/mL; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064