Retiro De Equipo (Recall) de BD ProbeTec¿ ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec¿ ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Becton Dickinson & Co..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    30844
  • Clase de Riesgo del Evento
    Class 1
  • Número del evento
    Z-0455-05
  • Fecha de inicio del evento
    2005-01-10
  • Fecha de publicación del evento
    2005-02-03
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2005-04-19
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Dna-Reagents, Neisseria - Product Code LSL
  • Causa
    An in-vitro diagnostic kit reagent may cause false negative clinical results in patients tested for sexually transmitted infections.
  • Acción
    BD notified consignees by phone on 01/10/05 and advised that users transporting patient urine samples may experience increased indeterminate results or false negative results when testing the samples on the BD ProbeTec¿ ET CT/GC Amplified DNA Assay. Subsequest to the phone notification, kit users identified as utilizing the Urine Processing Pouches were further notified by a letter. The letter describes 4 criteria contributing to the risk associated with false negative and indeterminate results. The letter advised users to discontinue use of the kits and discard for replacement. BD further advised in the notification that if the user is not utilizing the Amplification Control (AC), the firm recommends that they evaluate if urine specimens were transported and/or stored at refrigerated temperatures in the kit''s urine pouches. If specimens were transported and/or stored under the noted criteria, BD recommends that users should notify physicians who received reports of negative patient results and indicate that there is a possibility that these results were incorrect. BD plans to provide a 'Dear Physician' letter that would facilitate that notification upon request of the users.

Device

  • Modelo / Serial
    Lot/Exp 4103719 October 31, 2005 4118463 October 31, 2005 4119670 November 30, 2005 4124224 November 30, 2005 4133122 November 30, 2005 4105126 October 31, 2005 4118129 November 30, 2005 4104256 October 31, 2005 4091913 October 31, 2005 4104254 October 31, 2005 4090438 October 31, 2005 4090259 October 31, 2005 4084784 October 31, 2005 4014879 July 31, 2005 4016963 July 31, 2005 4023837 July 31, 2005 4023840 August 31, 2005 4041551 August 31, 2005 4083445 October 31, 2005 4133902 November 30, 2005
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Product was distributed to direct accounts in the US and to affiliated distributors internationally.
  • Descripción del producto
    BD ProbeTec¿ ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec¿ ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Becton Dickinson & Co., 7 Loveton Cir, Sparks MD 21152-9212
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA