Retiro De Equipo (Recall) de Bond ReadyToUse Primary Antibody Thyroid Transcription Factor1

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Leica Microsystems, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    56406
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2497-2010
  • Fecha de inicio del evento
    2010-08-02
  • Fecha de publicación del evento
    2010-09-25
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2012-01-29
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Immunohistochemistry Reagents And Kits - Product Code NJT
  • Causa
    In combination with a sensitive polymer detection system, the ttf-1 antibodies may stain additional tumor types other than is currently indicated on the instructions for use.
  • Acción
    Leica Microsystems sent Medical Device Field Correction letters dated 7/28/10 to the affected customers via first class mail on the same date, informing them that in combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor types other than is currently indicated on the Instructions for Use. Currently they state that lung and thyroid tumors may stain, but Leicas investigation further characterized the expression of the TTF-1 antigen and it may also be present in other tumors, e.g. thymomas. Customers who have received the affected lot numbers of the three products were requested to refer to the new Instructions for Use to review the amendments made to the Results Expected via the following links: http://bsd.leica-microsystems.com/pdfs/products/ttf-1-ce.pdf; http://bsd.leica-microsystems.com/pdfs/products/ttf-1-l-ce.pdf; and http://bsd.leica-microsystems.com/pdfs/products/pa0364.pdf; The accounts were asked to complete the attached Acknowledgement Form, acknowledging the receipt and understanding of the letter, listing the lot numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.

Device

  • Modelo / Serial
    Catalog #PA0364, lot numbers 05918, 06912, 07104, 08679, 09261, 10480, 10903
  • Clasificación del producto
  • Clase de dispositivo
    1
  • ¿Implante?
    No
  • Distribución
    Nationwide and Canada
  • Descripción del producto
    Bond Ready-To-Use Primary Antibody Thyroid Transcription Factor-1 (SPT24); 7 mL vial; an in vitro diagnostic reagent for the qualitative identification by light microscopy of human thyroid transcription factor-1 in formalin-fixed, paraffin- embedded tissue by immunohistochemical staining using the automated Bond system; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom; Catalog No.: PA0364