Retiro De Equipo (Recall) de Clintec Micromix Compounder

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Baxter Healthcare Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    28921
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1054-04
  • Fecha de inicio del evento
    2004-04-26
  • Fecha de publicación del evento
    2004-07-20
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2006-04-11
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel data.
  • Notas adicionales en la data
    unknown device name - Product Code NEP--
  • Causa
    The micromix compounder may have ingredient delivery outside of the stated accuracy limits if the vlier pin is not properly adjusted.
  • Acción
    Baxter began telephoning their customers on 4/26/04, advising them that an improperly adjusted Vlier pin could result in an ingredient delivery outside of stated accuracy limits. The accounts were requested to immediately stop using the Micromix Compounder and return the device to Baxter, and to evaluate all TPN formualtions that have been compounded but not administered to determine if any ingredients were delivered at a programmed volume of less than 1 mL. The account was advised to discard those solutions and recompound them using manual techniques. Follow-up recall letters were faxed to the accounts on the same date, reiterating the telephone calls. Baxter sent second recall letters dated 2/16/05 to the 15 customers who had not yet returned their Micromix Compounder as of 2/16/05, along with a copy of the 4/26/04 recall letter. The customers were informed of the new issue with the Micromix Compounder not recognizing decimal points for delivery volumes below 1 mL, and were again requested to return the Micromix Compounder and the handheld keypad to Baxter so that they may be corrected when the Micromix is reintroduced the market.

Device

  • Modelo / Serial
    product codes 2M8290 and 2M8290L, all serial numbers The recall extension concerns those units that had not been returned as of 2/16/05: serial numbes 10025, 11163, 11263, 11298, 11254, 11280, 10075, 11279, 11265, 11264, 10065, 11251, 11065, T12054, 10051
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide and internationally through Baxter subsidiaries to the United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan and Korea.
  • Descripción del producto
    Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA