Retiro De Equipo (Recall) de Colleague 3 Volumetric Infusion Pumps

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Baxter Healthcare Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    28824
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0880-04
  • Fecha de inicio del evento
    2004-04-13
  • Fecha de publicación del evento
    2004-07-20
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2011-02-02
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pump, Infusion - Product Code FRN
  • Causa
    Ink deterioration of the letters, numbers and symbols of the programming and pump channel keypads, which may impact the legibility of the keys over time.
  • Acción
    Baxter sent urgent device correction letters dated 4/13/04 to the direct accounts to the attention of the Director of Biomedical Engineering on the same date. The letters informed the accounts of the ink deterioration of the letters, numbers and symbols on the Colleague Volumetric Infusion Pump programming and pump channel keypads, and informed them that over time such ink deterioration may impact legibility of the keys. The letters provided a listing of model numbers and serial numbers of pumps at their facility which may have potentially affected keypads. The accounts were given a sample of the overlay to be applied over the programming keypad and installation instructions as an interim fix while replacement keypads are obtained and installed as the pumps are processed through their depot repair centers. The accounts were requested to complete the attached reply form and return it via fax to Baxter. The completed form acknowledges receipt of the recall letter, verifies the model numbers and serial numbers of the pumps at the facility and indicates the number of overlays needed. Any questions were directed to Baxter at 1-800-843-7867. On 6/9/05, Baxter sent Urgent Device Correction letters to the accounts, to the attention of the Director of Biomedical Engineering, advising them that the replacement pump programming and pump channel keypads were now available.

Device

  • Modelo / Serial
    product code 2M8153 - serial numbers 12110481CT and above; product code 2M8153K - serial numbers 12120001TK and above
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide, and internationally to Australia, the Bahamas, Belgium, Brazil, Canada, Colombia, El Salvador, Guatemala, Honduras, Israel, Lebanon, Mexico, New Zealand, Panama, the Philippines, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates and the United Kingdom.
  • Descripción del producto
    Colleague 3 Triple Channel Volumetric Infusion Pumps, product code 2M8153 and 2M8153K (United Kingdom); Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Singapore; monochrome display screen
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA