Retiro De Equipo (Recall) de Device Recall IGFBP3 ELISA Reagent Kit

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Beckman Coulter Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    58583
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2184-2011
  • Fecha de inicio del evento
    2011-03-30
  • Fecha de publicación del evento
    2011-05-11
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2011-12-02
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Radioimmunoassay, human growth hormone - Product Code CFL
  • Causa
    The recall was was initiated because patient results with the igfbp-3 elisa kit may be elevated by 70% compared to stated expected ranges in the instructions for use (ifu).
  • Acción
    Beckman Coulter sent a Product Corrective Action (PCA) letter with attached customer response form on March 30, 2011 to all customers who have the affected product, IGFBP- 3 Enzyme-Linked immunosorbent Assay ( ELISA) kit .This kit provides materials for the quantitative measurement of IGFBP-3 in serum. The letter informed the customers of the reason for recall, the products affected and actions to be taken. Consignees were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If they forwarded any of the affected product listed above to another laboratory, they were asked to provide a copy of the letter to them. Beckman further asked consignees to complete and return the enclosed Response Form within 10 days so they are assured consignees have received this important communication. Questions were directed to the Beckman Coulter Customer Support Center. - Website: http://www.beckmancoulter.com/customersupport/support - Phone, call 1-800-854-3633 in the United States and Canada - Outside the United States and Canada contact local Beckman Coulter representative.

Device

  • Modelo / Serial
    Lot Numbers: 992743, 992287, 991916, 991530, 991432, 990220 & 890830
  • Clasificación del producto
  • Clase de dispositivo
    1
  • ¿Implante?
    No
  • Distribución
    Worldwide -- USA, Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, France, Germany, Greece, Hong Kong, India, Italy, Korea, Republic of, Lebanon, Luxembourg, Morocco, Netherlands, New Zealand, Puerto Rico, Syrian Arab Republic, Taiwan, Thailand, Turkey & United Kingdom
  • Descripción del producto
    IGFBP-3 ELISA Reagent Kit; Part Number" DSL-10-6600; Lot Numbers: 992743, 992287, 991916, 991530, 991432, 990220 & 890830 || This assay is intended for in vitro diagnostic use.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Beckman Coulter Inc., 250 S Kraemer Blvd, Brea CA 92821-6232
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA