Retiro De Equipo (Recall) de Device Recall Integra External Fixation System

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Integra LifeSciences Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    74757
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2387-2016
  • Fecha de inicio del evento
    2016-07-20
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2016-12-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Appliance, fixation, nail/blade/plate combination, multiple component - Product Code KTT
  • Causa
    Through the investigation of reported complaints integra verified that there have been instances where the universal wire fixation bolts the slotted post broke during use at the surgical site.
  • Acción
    Integra sent an Urgent - Voluntary Medical Device Recall letter dated July 20, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Integra is asking you to take the following actions: 1. Please check your Integra¿ External Fixation System inventory to determine if you have the 17.5mm or 23mm Universal Wire Fixation Bolts and/or the large or small Slotted Posts. 2. If you do have 17.5mm or 23mm Universal Wire Fixation Bolts and/or large or small Slotted Posts, stop using them immediately. 3. Complete the attached "Acknowledgement and Return Form" and check the box: I do have affected products on the list and record the lot number. 4. Or, complete the attached "Acknowledgement and Return Form" and check the box: I do not have affected products. 5. Complete the other information as indicated on this form. Keep a copy of the form for your records. Return the completed "Acknowledgement and Return Form" by email or fax indicating your receipt and review of this notification. When your form is received, if you have noted you have affected products, an Integra Representative will contact you and provide you with directions to return the product, as well as input an order to replace the quantity you indicated on the form. Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement form for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that our customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact Sales Operations at 888-601-0203, option 2. We apologize for any inconvenience this may cause and thank you for your cooperation in this effort. Fo

Device

  • Modelo / Serial
    Lot # QJ0139, QJ0148, QJ0140, QJ0149, QJ0287
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    US Distribution to the states of : FL, VA, OK, CO, DC and IN.
  • Descripción del producto
    Integra External Fixation System Slotted Post - Small Catalogue #12224228; Integra External Fixation System Slotted Post Large Catalogue #12224229
  • Manufacturer

Manufacturer