Retiro De Equipo (Recall) de Device Recall Medtronic (TiMesh), Screwdriver, 1.5mm, Cruciate, Self Tapping Shaft

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Medtronic Neurosurgery.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    35046
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0849-06
  • Fecha de inicio del evento
    2006-03-08
  • Fecha de publicación del evento
    2006-05-06
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2008-06-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Screwdriver, Skullplate - Product Code GXL
  • Causa
    Medtronic neurosurgery has initiated this action because it was determined that two lots of the modular screwdriver blade components may exhiibt burrs along the tips of the screwdriver, which may prevent adequate engagement with the associated screws.
  • Acción
    Medtronic Neurosurgery sent a Sales Representative Notification letter to the applicable Medtronic Sales Representative by mail and fax on March 8, 2006. A Sales Representative Notification Letter was provided to each Medtronic Sales Representative who had received the product or is responsible for an account who has received the product. This letter informed the applicable Sales Representative of the action and instructed them how to complete the Sales Representative Customer Product Accountability Letter. A Sales Representative Customer Product Accountability Letter was provided to each Medtronic Sales Representative who has received the product or is responsible for an account who has received the product. This letter requests the current product status, i.e., has the product been used, discarded or will it be returned to Medtronic Neurosurgery. Each letter will be individualized for each customer indicating the specific catalog number(s), lot number(s) and quantity that were shipped to that customer. It is the firm''s intention to have the Sales Representative exchange all affected product with replacement product during this process.

Device

  • Modelo / Serial
    TV 59965,  TV 59967
  • Clasificación del producto
  • Clase de dispositivo
    1
  • ¿Implante?
    No
  • Distribución
    The product was distributed to the following states: Florida, Illinois, New York, Pennsylvania, Tennessee
  • Descripción del producto
    Medtronic (TiMesh), Screwdriver, 1.5mm, Cruciate, Self Tapping Shaft, Catalog Number 8001215 || The 8001215 Screwdriver is a stainless steel part designed for use with cruciate-style screws in the firm''s TiMesh System. The product is sold in non-sterile, single unit packages. They are re-usable, and autoclavable.