Retiro De Equipo (Recall) de Device Recall RS4i Sequential Stimulator;

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por International Rehabilitative Sciences, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    55655
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0323-2011
  • Fecha de inicio del evento
    2008-06-11
  • Fecha de publicación del evento
    2010-11-08
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2012-09-10
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Stimulator, muscle, powered - Product Code IPF
  • Causa
    The charge current is leaking through two diodes, which are supposed to block the reverse current, when the device is warm during the main battery charge operation, it then causes the lithium battery to swell around two years of use and it causes the device to be inoperable.
  • Acción
    Since June 11, 2008, RS Medical has taken the following actions when the device was sent back through a consumer complaint, or when the device was returned for routine maintenance, or through the rental service. The firm has upgraded the blocking diodes with higher temperature rated diodes, replaced the Lithium coin cell battery, and the software register bit has been changed to denote for subsequent automated testing that the diode upgrade/Lithium coin cell battery replacement has occurred. The firm reported that the device must be returned for the rechargeable NiMH battery replacement around two years of use, and the battery should not be replaced by patients. On 10/29/10, the firm began sending out URGENT NOTICE of VOLUNTARY RECALL letters to their customers. The firm will continue to send out the letters through 2011. The letter states the recall product as RS-4i Sequential Stimulator, Affected Serial Numbers 4i300000 to 4i467119. The customers are informed that "RS-4i Sequential Stimulators, only those within the affected serial number range indicated above, are susceptible to early expiration of the lithium coin cell battery that provides back-up power to the device memory when the stimulator is powered OFF. Customer experience indicates that the 3V lithium coin cell battery can be depleted as early as one third of its rated life. The RS-4i Stimulator will perform as intended until the lithium coin cell battery is fully depleted. When this happens, the Stimulator will turn OFF and fail to turn ON, even if it has been charged by battery charger connection to a standard electrical wall outlet". Customers are advised to make a schedule for a free Return and Repair of a RS-4i Sequential Stimulator and should call the Toll Free Service Line at 877-363-5753 (English) or 877-363-5754 (Spanish) from 6AM to 5PM (Pacific Time) Monday-Friday. If the customers have any other questions or to review additional information before scheduling a device

Device

  • Modelo / Serial
    Device is labeled as 4i and a unique sequential six-digit number. 4i is specific for RS 4i device and six-digit sequential numbering for device. Serial numbers 4i300000 to 4i467119.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide Distribution
  • Descripción del producto
    The device is labeled in part: " RS Medical RS-4i Stimulator". The device is sold in a kit that includes a carrying case, connector cables, pads (electrodes), battery charger, patient document, instructional DVD, return mail bag, and the operation manual. || This device is used with cables and electrodes that are placed on skin in the treatment area, and allow an electrical micro-current to relieve pain. The device requires a prescription from a physician. It is programmed by an RS medical employee (or Account Manager) to the physician's specific treatment plan for each patient. The patient is given instructions on how to utilize the device (proper electrode placement). Powered MUSCLE Stimulation indications for use including: 1) Relax muscle spasm. 2) Prevention or retardation of disuse atrophy. 3) Maintain or increase range of motion. 4) Increase local blood circulation. 5) Re-educate muscle. 6) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. INTERFERENTIAL Stimulation indications for use including: 1) Relieve acute pain. 2) Relieve and manage chronic pain.