Retiro De Equipo (Recall) de Device Recall Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit),

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Merge Healthcare, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    76946
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1923-2017
  • Fecha de inicio del evento
    2015-03-30
  • Fecha de publicación del evento
    2017-05-02
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2017-05-05
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Computer, diagnostic, programmable - Product Code DQK
  • Causa
    The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. this could result in the potential of a small bias added to the co2 measurement and affect the alarm notification.
  • Acción
    The recall was initiated by telephone on March 30, 2015. Follow-up phone calls were made July 6, 2015. The firm was unable to provide a telephone script. The recalling firm also issued letters dated March 28, 2016, via email on April 6, 2016, dated July 11, 2016. The letters contained the same content, notified the customer of the issue, and informed them a fix was available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they have the affected ISA CO2 sensors at their facility and are they interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. For further questions, please call (877) 741-5369.

Device

  • Modelo / Serial
    ISA CO2 Sensor serial numbers:  107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435,  107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776,  107831, 107832, 107836, 107839, and 107841.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
  • Descripción del producto
    Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.