Retiro De Equipo (Recall) de Device Recall Service parts for Crank Screw Assembly

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Medical Div. of Stryker Corporation.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    54265
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2343-2010
  • Fecha de inicio del evento
    2010-02-26
  • Fecha de publicación del evento
    2010-09-01
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2011-04-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Bed, manual - Product Code FNJ
  • Causa
    The fowler of the bed is raised by a crank. the crank consists of a steel screw that moves through a brass nut. over time the screw can wear the threads on the brass nut. the screw and the nut can become misaligned, causing a loud squealing noise when the screw comes in contact with the drive tube. in addition the fowler can become inoperable over time with excessive wear of the brass nut. t.
  • Acción
    The firm, Stryker Medical, sent an "Urgent Medical Device Correction" letter dated February 26, 2010, to all customers. The letter describes the product, problem and action to be taken by customers. The letter states that Stryker field service tech will contact the firm and perform device corrections within four months. The customers were instructed to locate the recalled stretchers and verify that there is no grinding noise when activating the fowler. If the fowler exhibited a grinding noise, the customers were instructed to remove the stretcher from service and contact the recalling firm at 800-STRYKER, option 3. Note: Do not put the stretcher back into service until it is repaired by our field service representative. The customers were also instructed to return the enclosed post card to confirm receipt of this notice; inform any new users and advise the firm of new locations if any of the listed stretchers were on loan or sold, and advise the firm of the obsolescence by providing the serial numbers if any of the listed stretchers were disposed of. If you have any urgent questions or concerns, contact (269) 324-6884. Normal hours are Monday-Friday 8am-5pm(EST).

Device

  • Modelo / Serial
    There is no code.
  • Clasificación del producto
  • Clase de dispositivo
    1
  • ¿Implante?
    No
  • Distribución
    Worldwide distribution: USA and countries including Australia, Canada, Chile, China, France, Germany, Greece, Spain, Italy, Japan, South Africa, Taiwan, Korea, Singapore, and UK.
  • Descripción del producto
    Service parts for Crank Screw Assembly (PN1069-042-010S and 1069-042110S) Stryker Medical 3800 E. Centre Ave Portage MI 49002. || Intended use: Service parts are replacement parts for the eye surgery stretcher.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Stryker Medical Div. of Stryker Corporation, 3800 E. Centre Ave., Portage MI 49002
  • Source
    USFDA