Retiro De Equipo (Recall) de Head Ring Assembly with Intubation Mounts (HRAIM)

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Integra LifeSciences Corporation.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    62978
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2429-2012
  • Fecha de inicio del evento
    2012-08-09
  • Fecha de publicación del evento
    2012-09-20
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2014-03-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Neurological stereotaxic Instrument - Product Code HAW
  • Causa
    T-handle screw is used on a complementary product to the crw, the integra hraim, which is the ct-compatible intubation head ring assembly. when the t-handle on the hraim intubation hoop is tightened, the t-handle could stop in a vertical position that prevents the crwprecise and certain models of the crw-asl from seating properly on the head ring although it appears to be fully seated.
  • Acción
    The firm, Integra, sent an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" dated August 10, 2012 to US & Canada consignees/customers via traceable courier service and Non-US & Non-Canada consignees/customers - through written notification delivered either by traceable courier, email, facsimile, or through a directed visit by an Integra Sales Representative. The notification describes the product, problem and actions to be taken. The Integra sales representative will replace the screws according to the attached rework instructions. If for some reason it is not possible to replace the screws during the visit, the sales representative will leave the screws and customers can then replace them by following the attached rework instructions and discard the old screws immediately after replacement. The customers were instructed to complete and return the Product Recall Acknowledgement Form via the sales representatives visit; fax at 1-609-750-7999 or email to: csnirecalls@integralife.com and include the number of devices to be identified and replaced . Note:The T -Handle screws that are replaced will be disposed of at the consignee's/customers location. Please feel free to contact our service hotline at 888-772-7378 should you have any additional questions.

Device

  • Modelo / Serial
    All lot numbers from 2002 to current
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.
  • Descripción del producto
    Integra HRAIM Head Ring Assembly with Intubation Mounts || The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CT/MR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes
  • Manufacturer

Manufacturer