Retiro De Equipo (Recall) de Omniflow 4000 Plus

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Hospira Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    52866
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0020-2010
  • Fecha de inicio del evento
    2009-08-11
  • Fecha de publicación del evento
    2009-10-06
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2010-11-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Notas adicionales en la data
    Infusion Pump - Product Code FRN
  • Causa
    Fire/shock hazard-- a component used in the device may cause sparks/flashes and poses a fire and shock hazard.
  • Acción
    Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.

Device

  • Modelo / Serial
    AC power cords Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010, and  826-96057-003.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution
  • Descripción del producto
    Omniflow 4000 Plus, List numbers 40051-04-05, 40051-04-07, 40051-04-09, 40051-04-51; AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Hospira Inc, 755 Jarvis Dr, Morgan Hill CA 95037-2810
  • Source
    USFDA