Retiro De Equipo (Recall) de Premisa

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Kerr Corp.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    49506
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0257-2009
  • Fecha de inicio del evento
    2008-09-05
  • Fecha de publicación del evento
    2008-11-25
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2012-03-07
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    dental composite - Product Code EBF
  • Causa
    Material stiff: material in several lots of premise unidose appears to stiffen and become difficult to extrude over time.
  • Acción
    Kerr sent an Urgent Medical Device Recall letter to consignees via USPS first class mail the week of September 8, 2008, requesting they return the product for replacement or credit. The recall notification requested the return of the affected products because the material appears to stiffen and become difficult to extrude over time. Customers were instructed if they have any of the affected lots of product in their inventory to immediately return it to Kerr. Any product returned will be replaced at no charge or credited to your account. In addition, customers were instructed to fax the attached Recall Return Form to 714-516-7739. If the customers were an authorized Kerr distributor, they were requested to identify and recover the affected product lots listed on the attached Lot Number Table that may have been shipped to their customers. Also, customers were instructed to contact Kerr Customer Care at 1 -800-537-7123 directly to handle the arrangements of a quick return and replacement

Device

  • Modelo / Serial
    LOT NUMBER(S): 3001316, and 3011566
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide: USA, Australia, Canada, Hong Kong, Malaysia, Mexico, and Singapore.
  • Descripción del producto
    PREMISE, PART NO. 32811, 10 PACK UNIDOSE PREMISA A1 (INTERNATIONAL ONLY), dental composite
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Kerr Corp, 1717 W Collins Ave, Orange CA 92867-5422
  • Source
    USFDA