Retiro De Equipo (Recall) de PROCEED Surgical Mesh is a sterile, thin, flexible laminate mesh designed for the repair of hernias and other fascial deficiencies. The mesh product is comprised of an oxidized regenerated cellulose (ORC) fabric, and PROLENE Soft Mesh, a nonabsorbable polypropylene mesh, which is encapsulated by a polydioxanone polymer. The polypropylene mesh side of the product allows for tissue ingrowth while the ORC side provides a bioresorbable layer that physically separates the polypropylene mesh from underlying tissue and organ surfaces during the wound-healing period to minimize tissue attachment to the mesh. The polydioxanone provides a bond to the ORC layer.

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Ethicon, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    34279
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0449-06
  • Fecha de inicio del evento
    2006-01-04
  • Fecha de publicación del evento
    2006-01-26
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2006-09-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Mesh, Surgical, Polymeric - Product Code FTL
  • Causa
    Lots of proceed surgical mesh may delaminate from the polypropylene mesh during certain hernia repairs. exposure of the polypropylene layer to the bowel prior to reperitonealization could increase the risk of adhesions and bowel fistulization.
  • Acción
    The recall was initiated on 12/21/2005. The recall to the customers were initiate via letter with return receipt on 1/4/2006.

Device

  • Modelo / Serial
    Product Code -- Lot: PCDD1---- UHG068; UJG096; UJG110; UJG131; UJG149; UKG264; ULG335; ULG316; ULG313; ULG342; ULG374; ULG385; ULG409; UMG490; UMG492; UMG004; UHG038; UHG025; UHG021;   PCDM1 ---- UJG078; UJG082; UJG085; UJG087; UJG109; ULG103; ULG101; ULG095; ULG125; UJG128; ULG139; UJG142; UKG184; UKG183; UKG181; UKG175; UKG161; UKG187; UKG203; UKG201; UEG883; UKG231; UKG239; UKG241; UKG279; UKG253;  UKG250; UKG243; UKG267; UKG266; UKG256; ULG309; ULG287; ULG328; ULG311; ULG370; ULG369; ULG340; ULG337; ULG393; ULG399; ULG400; ULG373; ULG401; ULG405; UMG424; UMG425; UMG426; UMG448; UMG449; UMG479; UMG478; UMG477; UMG473; UMG450; UMG504; UMG486; UGG001; UGG999; UHG014; UHG040; UHG031; UHG026;   PCDN1 ----- UHG056; UHG059; UHG072; UHG054; UJG075; UJG077; UJG088; UJG092; UJG099; UJG107; UJG134; UJG129; UJG122; UJG144; UJG154; UJG156; UJG160; UKG176; UKG170; UKG169; UKG166; UKG177; UKG190; UKG198; UKG215; UKG213; UKG220; UKG222; UKG225; UKG233; UKG248; UKG260; ULG308; ULG305; ULG289; ULG285; ULG281; ULG278; ULG272; ULG271; ULG269; ULG331; ULG321; ULG317; ULG315; ULG344; ULG343; ULG346; ULG352; ULG371; ULG368; ULG365; ULG361; ULG359; ULG341; ULG390; ULG392; ULG394; ULG396; UMG432; UMG427; ULG413; UMG438; UMG445; UMG472; UMG471; UMG459; UMG456; UMG455; UMG476; UMG493; UMG505; UMG509; UHG005; UHG019; UHG022; UHG034; UHG043; UHG044;    PCDL1 ----- UMG489; UMG483; UMG468; UMG451; UMG431; ULG408; UMG429; UMG430; ULG398; ULG397; ULG381; ULG339; ULG354; ULG355; ULG306; ULG319; ULG332; ULG193; ULG298; ULG279; ULG275; UKG255; UKG229; UKG227; UKG226; UKG168; UKG173; UKG178; UJG117; UJG094; UHG053; UHG069; UJG080; UHG060; UJG073; UHG048;   PCDH1 ----- UMG475; UMG481; UMG494; UMG436; UMG437; UMG454; UMG458; UMG457; UMG423; UMG433; UMG414; UMG415; ULG375; ULG376; ULG377; ULG391; ULG406; ULG372; ULG349; ULG348; ULG336; ULG327; ULG310; ULG304; ULG300; ULG288; ULG276; UKG265; UKG252; UKG218; UKG219; UKG230; UKG217; UKG205; UKG204; UKG186; UKG174; UKG167; UKG163; UJG157; UJG145; UJG143; UJG141; UJG127; UJG100; UJG098; UJG115; UJG114; UJG108; UJG084; UGG997; UGG998; UHG013; UHG030; UHG039;    PCDB1 ------ ULG323; ULG296; UEG886; UEG884; UJG116; UJG108; UJG104; UJG089; UJG079; UHG065; UJG076; UHG067; ULG296; UEG884; UEG886; UJG116; UJG104; UJG106; UJG089; UJG079; UHG065; UJG076; UHG067; ULG296; UEG884; UEG886; UJG116; UJG104; UJG106; UJG089; UJG079; UHG065; UJG076; UHG067; ULG323; ULG296; UEG884; UEG886; UJG116; UJG104; UJG106; UJG079; UJG089; UHG065; UJG076; UHG067;  PCDG1 ----- UMG482; UMG498; UMG499; UMG500; UMG441; UMG469; UMG470; UMG442; UMG447; UMG446; ULG402; UMG416; UMG419; UMG420; UMG421; UMG422; ULG384; ULG386; ULG387; ULG395; ULG356; ULG366; ULG367; ULG338; ULG329; ULG326; ULG325; ULG274; ULG334; ULG307; ULG299; ULG294; ULG288; ULG284; ULG283; UKG258; UKG259; UKG263; ULG277; UEG885; UKG234; UKG244; UKG245; UKG254; UKG196; UKG210; UKG197; UKG191; UKG193; UKG162; UKG179; UKG180; UJG146; UJG148; UJG153; UJG159; UJG132; UJG130; UJG124; UJG121; UJG105; UJG113; UJG091; UJG086; UJG083; UHG012; UHG063; UHG061; UHG047; UHG071; UHG064; UHG058; UHG055; UHG002; UHG003; UHG011; UHG017; UHG027; UHG036; UHG037; UHG041; UHG042; UHG045;   PCDT1 ----- UMG488; UMG485; UMG491; UMG452; UMG453; UMG435; ULG407; ULG410; UMG417; UMG428; UMG434; ULG363; ULG364; ULG389; ULG345; ULG357; ULG358; ULG291; ULG273; UKG257; UKG236; UKG221; UKG208; UKG209; UKG192; UKG171; UJG137; UJG111; UJG074; UHG057; UHG051; UHG020; UHG032; UHG033; UHG049;   PCDR1 ----- UMG487; UMG502; UMG474; UMG480; ULG333; ULG347; ULG301; ULG290; ULG282; UKG228; UKG237; UKG240; UKG195; UKG172; UJG147; UJG150; UJG140; UJG138; UJG136; UJG123; UJG102; UJG093; UJG090; UHG070; UHG062; UHG046; UHG052; UGG996; UHG006; UHG008; UHG009; UHG010; UHG016; UHG028;   PCDW1 ----- UMG495; UMG496; UMG444; UMG443; ULG412; UMG418; ULG378; ULG388; ULG380; ULG379; ULG360; ULG320; ULG297; ULG292; ULG280; UKG261; UKG251; UKG247; UKG189; UEG887; UKG206; UKG202; UKG188; UJG120; UJG151; UJG133; UHG029; UHG050;   PCDJ1 ------- UMG484; UMG497; UMG439; UMG440; ULG383; ULG382; ULG403; ULG404; ULG362; ULG353; ULG350; ULG351; ULG330; ULG314; ULG312; UKG238; ULG303; ULG318; ULG270; ULG268; UKG262; UKG246; UKG235; UKG232; UKG199; UKG212; UKG211; UKG200; UKG199; UKG185; UKG194; UKG182; UKG165; UKG164; UJG155; UJG152; UJG119; UJG165; UJG118; UJG112; UJG097; UJG081; UHG066; UGG990; UHG007; UHG015; UHG018;
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    There is a total of 1,070 consignees that have received recall product divided between hospitals and distributors nationwide. The product has been distributed internationally to the following countries: Argentina, Australia, Austra, Brazil, Canada, Chile Colombia, Czechoslovakian Republic, France, Germany Greece, Guaynabo, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Republic of Yemen, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirate, United Kingdom, Venezuela. Government shipments include: Alaska Area Native Health, Anchorage, AK; Audie L. Murphy VA Hospital, San Antonio, TX; VA Warehouse, Sioux Falls, SD; VA Medical Center, Miami, FL; Edward Hines Jr. VA Hospital, Hines, IL; J.L. McClellan VA Med. Ctr, Little Rock, AR; Medical Supply Warehouse, Andrews AFB, MD; Mike O''Callaghan Fed. Hospit, Neillis AFB, NV; Naval Medical Ctr., Portsmouth, VA; Recv Officer Hanal Hosp, Jacksonville, FL; USAF Academy Hospital, USAF Academy, CO; USAF Hospital Elmendorf, Elmendorf AFB, AK; VA Bay Pines, Saint Petersburg, FL; VA Hosp. Supply Service, Dallas, TX; VA Louisville, Louisville, KY; VA Med. Center WTS, Indianapolis, IN; VA Med. Ctr., Huntington, WV; VA Med. Ctr., Los Angeles, CA; VA Med. Ctr., Buffalo, NY; VA Med. Ctr., San Juan, PR; VA Med. Ctr, East Orange, NJ; VA Med. Ctr., New York, NY; VA Med. Ctr., Kansas City, MO; VA Med. Ctr., Leavenworth, KS; VA Med. Ctr., Wichita, KS; VA Med. Ctr., Temple, TX; VA Med. Ctr., Amarillo, TX; VA Med. Ctr., Grand Junction, CO; VA Med. Ctr., Nashville, TN; VA Med. Ctr., Mountain Home, TN; VA Med. Ctr., Loma Linda, CA; VA Med. Ctr., Wilmington, DE; VA Med. Ctr., Salisbury, NC; VA Med. Ctr., Columbia, MO; VA Med. Ctr., Saint Louis, MO; VA New York Harbor Healthcare, Brooklyn, NY; WHMC/MSLS, Lackland AFB, TX; Womack Army, Fort Bragg, NC.
  • Descripción del producto
    PROCEED Surgical Mesh is a sterile, thin, flexible laminate mesh designed for the repair of hernias and other fascial deficiencies. The mesh product is comprised of an oxidized regenerated cellulose (ORC) fabric, and PROLENE Soft Mesh, a nonabsorbable polypropylene mesh, which is encapsulated by a polydioxanone polymer. The polypropylene mesh side of the product allows for tissue ingrowth while the ORC side provides a bioresorbable layer that physically separates the polypropylene mesh from underlying tissue and organ surfaces during the wound-healing period to minimize tissue attachment to the mesh. The polydioxanone provides a bond to the ORC layer.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Ethicon, Inc., US Highway 22 West, Somerville NJ 08876