Retiro De Equipo (Recall) de Satellite (Medtronic Sofamor Danek) Satellite Sphere CoCr

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Medtronic Sofamor Danek USA Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    44967
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0199-2008
  • Fecha de inicio del evento
    2007-09-13
  • Fecha de publicación del evento
    2007-12-22
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2009-08-18
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Spinal System - Product Code NVR
  • Causa
    Labeling, including indications for use, lack 510 (k) approval.
  • Acción
    On September 12, 2007, the recalling firm's representatives were instructed by electronic mail to remove any manuals in the field and contact surgeons who have evaluated, implanted or been trained to implant the Satellite System and to collect any guides they may have on site. The sales representatives that have accounts where the Satellite product was shipped directly were mailed the information and instructed to hand deliver the 9/13/07, Urgent Device Recall notices to each consignees risk management office. The recall was then expanded on 10/31/2007, to include the surgical manual, the implants and associated instruments. The delivery of the second Urgent Device Recall Notices, dated 10/25/07, were handled in three ways: 1) The letter was delivered to all initial consignees that received satellite products along with a questionnaire (response card) to be completed by the Risk Manager or equivalent administrative function., 2) The letter without the questionnaire was sent directly to target surgeons who are known to have implanted or have been trained to implant Satellite Product 3) a field action confirmation form was sent to Sales Representative with surgeon customers that are receiving the recall communication. The form will be completed by the Sales Representative to confirm the surgeons have received the recall communication and have returned any of the surgical techniques and instruments.

Device

  • Modelo / Serial
    ALL LOTS to include: W05K0561, W05K0582, W05K1920, W06A2074, W07H0210
  • Clasificación del producto
  • Clase de dispositivo
    Unclassified
  • ¿Implante?
    Yes
  • Distribución
    Nationwide
  • Descripción del producto
    Medtronic Sofamor Danek, 14 mm Satellite Sphere CoCr, Size 14 mm, Quantity 1 each, Material Cobalt Chrome, Sterile, Rx only, REF 8000214, Manufactured at: Warsaw, Indiana, USA
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Medtronic Sofamor Danek USA Inc, 1800 Pyramid Place, Memphis TN 38132-1719
  • Source
    USFDA