Retiro De Equipo (Recall) de Spine Pack

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Windstone Medical Packaging, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    71968
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0271-2017
  • Fecha de inicio del evento
    2015-04-22
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2016-10-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Light, surgical, ceiling mounted - Product Code FSY
  • Causa
    The medtronic cover light handle disposable gloves, part # 13140, of the custom ams kit were recalled because gloves may contain splits or holes compromising sterility.
  • Acción
    Aligned Medical Solutions (AMS) sent the Medtronic Urgent Medical Device for Devon Light Glove letter, dated April 23, 2015 to all affected consignees. Consignees were informed that the sterile kits assembled and distributed by AMS contain the recalled Devon COVER LIGHT HANDLE DISPOSABLE Glove (AMS part #13140 or Medtronic part # 571711) due to splits or holes. Consignees were advised to identify any remaining affected kits and follow the over-labeling instruction. At the time the kit is opened for use, any light gloves should be identified and disposed of pursuant to the medical waste policies in effect at your institution. AMS can provide replacement product if needed or you can contact your local Medtronic sales rep for sterile replacements as well. Due to the manufacturing process and finished sterilized packs in inventory, or at your facility, over labeling of inventory and/or future shipment affected by this recall will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE Medtronic announced the recall l of Devon Light Glove contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. 1. At the time the kit is opened for use any Light Glove manufactured by Medtronic should be identified and set aside 2. The recalled light glove should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Consignees should complete the attached Reply Form and fax to AMS at 406-256-9875. For any questions about this recall, please call 321-527-7714. For questions regarding this recall call 404-259-6387.

Device

  • Modelo / Serial
    Lot Number/ Expiration Date:  75946 6/22/2016 76964 6/30/2016 76134 7/23/2016 76981 8/12/2016 77637 10/31/2016
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide Distribution to California, Florida, Georgia, Illinois, Kansas, Louisiana, Michigan, New Jersey, Nevada, Texas, Virginia, and Wyoming.
  • Descripción del producto
    Spine Pack, Kit number AMS3753 and AMS3783(A || convenience custom kits used for general surgery in hospital operating room
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Windstone Medical Packaging, Inc., 1602 4th Ave N, Billings MT 59101-1521
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA