Retiro De Equipo (Recall) de STERRAD 100NX Cassettes

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Advanced Sterilization Products.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    66317
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0044-2014
  • Fecha de inicio del evento
    2013-09-18
  • Fecha de publicación del evento
    2013-10-22
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2014-01-13
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Sterilizer, chemical - Product Code MLR
  • Causa
    Advanced sterilization products (asp) is recalling the three lots (13f049, 13f050, 13f051) sterrad 100nx cassettes because asp has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
  • Acción
    Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.

Device

  • Modelo / Serial
    Lot# 13F050, 13F051, 13F049
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide Distribution
  • Descripción del producto
    STERRAD 100NX Cassettes || The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Advanced Sterilization Products, 33 Technology Dr, Irvine CA 92618-2346
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA